Error Messages; External Parameters Setting; External Calibration - Baxtran PD6LED Manuel D'usager

Bascule électronique
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

User's Guide
Indicate that the scale is overloaded
- - - - -
Indicate that the voltage of the batteries is low, and the scale can't work normally
Lo
Indicate that the weight of the objects on the tray is outside the zero-resetting range
oUtZ
Indicate that the calibration is failed
Err1
Press and hold
parameters.
Automatic power-off function setting
off 0:
Shut up the automatic power-off function.
off 1:
2 minutes automatic power-off.
off 2:
5 minutes automatic power-off.
Use
key to choose and use
Energy-saving function setting (only apply to LED display)
EC 0:
Energy-saving function is closed
EC 1:
Energy-saving function is open
Backlight function setting (only apply to LCD display)
Backlight function is closed
BL 0:
BL 1:
Backlight function is open
Use
key to choose and use
Beep status setting
bp 0:
Whether stable or not, the buzzer beep
When stable, the buzzer beep
bp 1:
Use
key to choose and use
1. Press and hold
2. Remove all objects from the tray, press again, the screen will show "0 0 0 0". Press
the twinkling number. Press
3. Place a weight which value equal to the number on the tray (unit: g) (eg. the shown number is 6000,
that means you should place a 6000g weight on the tray) and then press
If success it will return to normal weighing.
NOTE: If after calibration your scale does not read accurately, it means the calibration is failed and the
calibration process should be repeated more slowly.
Please calibrate on a very stable and flat surface!
16
key, then press
key to turn on the scale, it will enter to setting the external
key to confirm.
key to confirm.
key to confirm. Then return to normal weighing.
key, then press
key to turn the scale, it will show "CAL".
key to increase the value of the twinkling number.

5. ERROR MESSAGES

6. EXTERNAL PARAMETERS SETTING

7. EXTERNAL CALIBRATION

PD model
key to shift
key to confirm.