Mains Part - Beurer BM 55 Mode D'emploi

Tensiomètre
Masquer les pouces Voir aussi pour BM 55:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 36
Classification
Internal supply, IPX0, no AP or APG,
continuous operation, application part
type BF
The serial number is located on the device or in the battery
compartment.
Technical information is subject to change without notification
to allow for updates.
• This unit is in line with European Standard EN 60601-1-2 (In
accordance with CISPR 11, IEC 61000-4-2, IEC 61000-4-3,
IEC 61000-4-4, IEC 61000-4-5, IEC 61000-4-6, IEC 61000-
4-8, IEC 61000-4-11) and is subject to particular precautions
with regard to electromagnetic compatibility (EMC). Please
note that portable and mobile HF communication systems
may interfere with this unit.
• This device is in line with the EU Medical Devices Directive
93/42/EEC, the "Medizinproduktegesetz" (German Medical
Devices Act) and the standards EN 1060-1 (non-invasive
sphygmomanometers, Part 1: General requirements),
EN 1060-3 (non-invasive sphygmomanometers, Part 3:
Supplementary requirements for electro-mechanical blood
pressure measuring systems) and IEC 80601-2-30 (Medical
electrical equipment – Part 2 – 30: Particular requirements
for the safety and essential performance of automated non-
invasive blood pressure monitors).
• The accuracy of this blood pressure monitor has been care-
fully checked and developed with regard to a long useful
life. If using the device for commercial medical purposes, it
must be regularly tested for accuracy by appropriate means.
Precise instructions for checking accuracy may be requested
from the service address.

12. Mains part

Model no.
LXCP12-006060BEH
Input
100 – 240 V, 50 – 60 Hz, 0.5 A max
Output
6 V DC, 600 mA, only in connection with
beurer blood pressure monitor.
Supplier
Shenzhen Iongxc power supply co., ltd
Protection
This device is double insulated and
protected against short circuit and
overload by a primary thermal fuse.
Make sure to take the batteries out
of the compartment before using the
mains adapter.
Polarity of the the DC voltage
connection
Double insulated / equipment class 2
Enclosures and
Equipment enclosed to protect against
Protective Covers
contact with live parts, and with parts
which can become live (finger, pin,
hook test).
The operator shall not contact the pa-
tient and the output plug of AC mains
adapter simultaneously.
34

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Bm-55M1002

Table des Matières