Make sure that the cuff tube is properly inserted and that you
do not move or talk.
Re-insert the batteries if necessary, or else replace them.
Technical alarm – description
Should the recorded blood pressure (systolic or diastolic) lie
outside the limits specified in the section "Technical specifica-
tions", the technical alarm will appear on the display indicating
either "     " or " Lo  " . In such cases, you should seek medical as-
sistance and check the accuracy of your procedure.
The limit values for the technical alarm are factory set and can-
not be adjusted or deactivated. These alarm limit values are
accorded second priority under the standard IEC 60601-1-8.
The technical alarm is a non-locking alarm and must not be
reset. The signal shown on the display will disappear automati-
cally after about 8 seconds.

11. Specifications

Model no.
Type
Measurement
method
Measurement range
Display accuracy
Measurement
inaccuracy
Memory
Dimensions
Weight
Cuff size
Permissible operating
conditions
Permissible storage
and transport condi-
tions
Power supply
Battery life
33
BM 55
M1002
Oscillometric, non-invasive blood pres-
sure measurement on the upper arm
Cuff pressure 0 – 300 mmHg,
systolic 60– 260 mmHg,
diastolic 40 –199 mmHg,
Pulse 40 –180 beats/minute
Systolic ± 3 mmHg, diastolic ± 3 mmHg,
pulse ± 5 % of the value shown
Max. permissible standard deviation
according to clinical testing:
systolic 8 mmHg /diastolic 8 mmHg
2 x 60 memory spaces
L 186 mm x W 95 mm x H 56 mm
Approx. 467 g (without batteries)
22 to 42 cm
+10 °C to + 40 °C, ≤ 90 % relative air
humidity (non-condensing)
- 20 °C to + 55 °C, ≤ 90 % relative air hu-
midity, 800 –1050 hPa ambient pressure
4 x 1.5 V
AAA batteries
For approx. 200 measurements, de-
pending on the blood pressure level
and/or pump pressure