Siemens SITRANS F Serie Guide Rapide page 10

Masquer les pouces Voir aussi pour SITRANS F Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Introduction
1.2 Safety Notes
Flow Meter Markings and Explanations
● IP65 – Ingress protection against solid bodies, rating of dust-tight and against liquid,
Transducers Markings and Explanations
● IP65 – Ingress protection against solid bodies, rating of dust-tight and against liquid,
Quick Start Guide Safety Information for Hazardous Areas
Note
Ratings under this heading apply to specific model families
Check Your Model Number:
FUS1010 7ME3532, 7ME3533, FUH1010 7ME3602, 7ME3603, FUG1010, 7ME3612 and
7ME3613 only;
FM-CSA installation
Read, understand and follow all safety instruction on the electronic media provided. This
equipment is rated for use in hazardous (classified) locations as stated below and must be
installed according to the 1010-443 installation drawing provided on the media. Failure to
install the equipment in the prescribed manner will result in unsafe operation. Follow all local
jurisdictional safety codes when operating this equipment. When properly installed the
equipment meets the following FM – CSA ratings:
Flow Meter
● Explosionproof for Class I, Division1, Groups B, C, D;
● Dust-ignitionproof for Class II, Division 1, Groups E, F and G
● Intrinsically safe connections for Class I and II, Division 1, Groups A, B, C, D, E, F and G;
● Nonincendive for Class I, Division 2, Groups A, B, C and D;
● Suitable for Class II Division 2 Groups F and G outdoor (Type 4X)
8
II (1) G [EEx ia] IIC – Flow meter located in the non-hazardous area with intrinsically
safe circuits of category Ex ia, which can be connected to Category 1 transducers
II 3 (1) G EEx nC [ia] IIC T5 – Category 3 Flow meter located in Zone 2 hazardous
area with intrinsically safe circuits of category Ex ia, which can be connected to Category
1 transducers in Zone 0
rating of water jets
II 1 G EEx ia IIC T5 – Category 1 Transducers located in Zone 1 hazardous area with
intrinsically safe circuits of category Ex ia for use in potentially explosive atmosphere
containing gases
rating of water jets
SITRANS FUH1010 Quick Start Guide
Operating Instructions, 1/2009, CQO:QSG009 Rev A

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Sitrans fuh1010

Table des Matières