13.3 Printer operation
Printout examples:
Prt
0 / 2
1 / 3
14 Error messages
Display
Description
Zero range exceeded
(on start-up or when pressing the
• Load on weighing pan
• Excess load, during zero setting of weighing scale
• Incorrect adjusting process
• Fault on load cell
Value outside the A/D converter range
• Damaged weighing cell
• Damaged electronics
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
40
60.0kg
60.0kg
170.0cm
20.7BMI
key)
MTA-M-BA-e-1612