8 Electromagnetic compatibility (EMC)

8.1 General hints

The installation and use of the electrical medical personal weighing
scales MTA-M requires special precautionary measures as outlined in
the EMC information below.
This device complies with the limits set for medical electrical devices of group 1,
class B (as per EN 60601-1-2).
Electromagnetic compatibility (EMC) describes a device's ability to perform reliably
within an electromagnetic environment without causing inadmissible electromagnetic
interference at the same time. Amongst other things, such disturbances may be
emitted by connecting cables or the air.
Inadmissible disturbances from the environment may result in incorrect displays,
inaccurate measured values or incorrect behaviour of the personal weighing scales
MTA-M. By the same token the personal weighing scales MTA-M may in some cases
cause such disturbances in other devices. To eliminate problems of that kind, we
recommend you to take one or several of the measures listed below:
• Change the alignment or distance of the device to the source of EMI.
• Install or use the personal weighing scales MTA-M at a different location.
• Connect the personal weighing scales MTA-M to a different power source.
• For further questions please contact our customer services.
Disturbances may be caused by improper modification or add-ons to the device or
not recommended accessories (such as power units or connecting cables). The
manufacturer will not be responsible for these. Modifications may also result in a loss
of authorisation relating to the use of the device.
Devices emitting high frequency signals (mobile telephones, radio
transmitters, radio receivers) may cause interference in the personal
weighing scales MTA-M. For that reason do not use them near the
personal weighing scales MTA-M. Chapter 8.4 contains details about
recommended minimum distances.