Material Resistance - Amann Girrbach artex gesichtsbogen Instructions

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

- sterilization time (exposure time at the sterilization
temperature):
Area
fractionated
vacuum/dynamic
air removal
USA
at least 4 min at
132 °C (270 °F),
drying time at
least 20 min
other
at least 3 min
countries
132 °C (270
°F)/134 °C (273
°F)
The fundamental suitability of the components for
an effective steam sterilization was demonstrated
by an independent accredited and recognized (§
15 (5) MPG) test laboratory by application of the
steam sterilizer HST 6x6x6 (Zirbus technology
GmbH, Bad Grund) and the fractionated vacuum/
dynamic air removal procedure as well the gravity
displacement procedure. For this, typical condi-
tions in clinic and doctor´s practice as well as the
specified procedure were considered.
The flash/immediate use sterilization procedure must
not be used.
Do not use dry heat sterilization, radiation steriliza-
tion, formaldehyde and ethylene oxide sterilization,
as well as plasma sterilization.
STORAGE
Please store the components after sterilization in the
sterilization packagings at a dry and dust-free place.
3 at least three vacuum steps
4 The less effective gravity displacement procedure must not be used in case of availability of the fractionated vacuum procedure.
5 The effectively required drying time depends directly on parameters in sole responsibility of the user (load configuration and density, sterilizer condi-
tions, ...) and by this is to be determined by the user. Nevertheless, drying times less than 20 min must not be applied.
6 respectively 18 min (inactivation of prions, not relevant for USA)
I N S T R U C T I O N S F O R C L E A N I N G A N D C A R E
I N S T R U C T I O N S F O R P R O C E S S I N G
gravity
displacement
at least 30 min
at 121 °C (250 °F),
drying time at
least 20 min
5
5
at
at least 20 min
6
at 121 °C (250 °F)

MATERIAL RESISTANCE

Please take care that the listed substances are not in-
gredients of the cleaning or disinfection detergent:
- organic, mineral, and oxidizing acids (minimum ad-
mitted pH-value 6.5)
- lyes (maximum admitted pH-value 8.5, neutral/en-
zymatic cleaner recommended)
- organic solvents (for example: acetone, ether, alco-
hol, benzine)
- oxidizing agents (for example: peroxide)
- halogens (chlorine, iodine, bromine)
- aromatic, halogenated hydrocarbons
Please do not clean any components by use of metal
brushes or steel wool.
Please do not applied acidic neutralization agents or
any rinse aids.
Please do not expose any components to tempera-
tures higher than 142 °C (288 °F)!
REUSABILITY
The components can be reused – in case of adequate
care and if they are undamaged and clean – 50times.
The user is responsible for each further use as well
as for the use of damaged and dirty components (no
liability in case of disregard).