Performance Approvals
Response time
t
0...20
0-5 Vol% O
2
≤ 14 seconds
0-25 Vol% O
≤ 10 seconds
2
1) Response time is higher than the permissible limit of EN 50104.
Stabilization time: 5 x t
0...90
Measuring range
0-5...25 Vol% O
, 0-25 Vol% O
2
Default: 25 Vol% O
2
Minimum reading: -1.25 Vol% O
Warm-up time
Operation: ≤ 20 minutes / Calibration: ≤ 6 hours
19.4 FM Performance Approval for
DrägerSensor H
The DrägerSensor H
S and H
2
diffusion sensors measuring hydrogen sulfide (H
ambient air.
DrägerSensor H
S (6810435)
2
Examination Type P8100
1)
Interface
Environmental parameters
Humidity
Temperature
Storage temperature
2)
Enclosure protection
Range
Accuracy (whichever is
greater)
Accessories
Dust filter
Adapter kit
Standard
1) HART communication shall not be used for safety-relevant communi-
cation
2) After submersion and jet water the sensor shall be replaced
DrägerSensor H
S LC (6809610)
2
Examination Type P8100
1)
Interface
Environmental parameters
Humidity
Temperature
40
t
0...90
1)
≤ 45 seconds
≤ 27 seconds
2
2
S and H
S LC
2
2
S LC are electrochemical
2
S) in
2
ETR 0400, 0401, 0500, 0501
4-20mA, relay
5 to 95%RH, non-condensing
-40/+65°C (-40/149°F)
0/+40°C (32/104°F)
IP 65/66/67
0-100ppm
measurement uncertainly (of
meas. value) ≤ ±3 % or mini-
mum ≤ ±0.5ppm
6812223
6810536
ANSI/ISA-92.00.01FM6340
ETR 0400, 0401, 0500, 0501
4-20mA, relay
5 to 95%RH, non-condensing
-20/+65°C (-4/149°F)
DrägerSensor H
S LC (6809610)
2
Storage temperature
2)
Enclosure protection
Range
Accuracy (whichever is
greater)
Accessories
Dust filter
Adapter kit
Standard
1) HART communication shall not be used for safety-relevant communi-
cation
2) After submersion and jet water the sensor shall be replaced
Instructions for use
0/+40°C (32/104°F)
IP 65/66/67
0-100ppm
measurement uncertainly (of
meas.value) ≤ ±3 % or mini-
mum ≤ ±0.5ppm in the range
>50ppm, ±10 % or ±3ppm
6812223
6810536
ANSI/ISA-92.00.01FM6340
®
|