Drägersensor Ols (6809630) - Dräger Polytron 8000 Serie Notice D'utilisation

Masquer les pouces Voir aussi pour Polytron 8000 Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Storage
Time
Influence of environmental parameters
Zero point
2)3)
Temperature
-20 to 55 °C
< ± 0.2 Vol%O
-10 to 55 °C
-
-10 to -20 °C
-
Pressure
< ±0.2 Vol%O
Humidity
No influence
1) Relative deviation of display reading is caused by atmospheric pres-
sure changes.
2) At temperatures below -5 °C the measurement deviation is higher
than specified in the EN 50104.
3) For operation-temperatures outside -5 °C to 40 °C a calibration at
operation temperature must be performed.
1)
Response time
t
0...20
0-5 Vol% O
2
≤ 15 seconds
0-25 Vol% O
≤ 10 seconds
2
1) At temperatures below -5 °C, the response time may increase.
2) Response time is higher than the permissible limit of EN 50104.
Stabilization time: 5 x t
0...90
Measuring range
0-5 Vol% O
to 0-100 Vol% O
2
Default: 25 Vol% O
Minimum reading: -1.25 Vol% O
2
Warm-up time
Operation: <20 minutes / Calibration: ≤ 2 hours
Instructions for use
|
Dräger Polytron
Storing of sensors is not
intended. Sensors should be
used upon arrival. Remaining
service life = Expected Service
life - Storage time
1)
Span
Relative deviation
2
from reading at
20 °C
< ± 8%
< ± 16%
Relative deviation
2
from reading at
1013 hPa:
< 10% of measure-
ment value /
100 hPa
Relative deviation
from reading at
50 % r.H.:
< 1 % of measure-
ment value
t
0...90
2)
≤ 33 seconds
≤ 26 seconds
2
2
®
8000 Series
19.3.2 DrägerSensor O
The DrägerSensor O
LS (6809630) is an electrochemical 3-
2
electrode sensor for measuring Oxygen (O
Environmental parameters
Pressure
Humidity
Temperature
Storage
Pressure
Humidity
Temperature
Time
Influence of environmental parameters
Zero point
Temperature
-40 to 65 °C
< ± 0.3 Vol%O
Pressure
< ± 0.1 Vol% O
Humidity
No influence
When measuring oxygen in the presence of helium, helium
compensation can be used. The option of helium
compensation is no certified measuring function according to
EN 50104.
CAUTION
Sensor failure
The exposure to higher concentrations of unsaturated
hydrocarbons, alcohols or hydrogen over an extended period
of time (dose approx. 100,000 ppm x hours) could cause the
sensor to fail.
► Ensure that the sensor is not exposed to such
concentrations over an extended period of time.
Performance Approvals
LS (6809630)
2
) in ambient air.
2
20.7 to 38.4 in. Hg (700 to
1300 hPa)
5 to 95%RH, non-condensing
-40 to +60 °C
short term +65 °C
no influence
30 to 70%RH, non-condensing
(only relevant if sensor packag-
ing open)
0 to +40 °C
Storing of sensors is not
intended. Sensors should be
used upon arrival. Remaining
service life = Expected Service
life - Storage time
Span
< ± 0.3 Vol%O
2
2
Relative deviation
2
from reading at
1013 hPa:
< 2 % of measure-
ment value /
100 hPa
Relative deviation
from reading at 50 %
r.H.:
< 2.5 % of measure-
ment value
|
en
39