Troubleshooting
Fault
Cause
number
002,
Serious data error
025 - 027,
in the instrument,
030 - 034,
various causes.
041, 042,
044, 050,
052, 081,
085
010
4 to 20 mA inter-
face cable not con-
nected.
045
Instrument not
detecting any sen-
sor.
051, 054
Faulty zero calibra-
tion.
055
Faulty span cali-
bration.
076,
Instrument fault.
080 - 084
090
Sensor lock func-
tion is activated. A
sensor with a dif-
ferent part number
has been con-
nected.
105
Instrument fault.
137, 64,
Instrument fault.
91
®
Polytron
87x0 IR-specific:
Fault
Cause
number
064, 071
Communication
fault.
083
PIR 7x00 optics
dirty.
086
Supply voltage out-
side range
087
4 to 20 mA fault at
PIR 7x00 sensor.
094, 095
Data error in the
instrument.
22
Remedy
Reset the instrument to fac-
tory default settings. If this
fault occurs again: Have
the instrument checked by
DrägerService.
Check the 4 to 20 mA inter-
face connection.
If operated as a stand-
alone instrument without a
central controller, wire pin 3
to pin 4.
Check connections.If this
fault occurs again: Have
the instrument checked by
DrägerService.
Perform zero calibration.
Perform span calibration
Check electrical connec-
tions.If this fault occurs
again: Have the instrument
checked by DrägerService.
Use a sensor with the same
part number or disable the
sensor lock function.
Have the instrument
checked by DrägerService.
Cycle power.If this fault
occurs again: Have the
instrument checked by
DrägerService.
Remedy
Check connection to PIR
7x00.
Clean PIR 7x00 optics.
Check supply voltage.
Check 4 to 20 mA connec-
tion to PIR 7x00.
Reset the sensor to factory
default settings. If this fault
occurs again: Have the
instrument checked by
DrägerService.
Fault
Cause
number
096
SIL password does
not match with PIR
7x00.

10.2 Warnings

Warning
Cause
number
011
Supply voltage is
out of range.
101
Data logger in
Stack mode is 100
% full and is not
logging any more
data.
102
Data logger in
Stack mode is 90 %
full.
103, 106
Data error in the
instrument. Some
dongle functions
like data logger,
sensor test, etc.
may no longer be
available.
104
Date or time setting
not valid.
105
Instrument fault.
107
Battery for data
memory empty.
110, 111,
SW dongle has
112
been removed
without being deac-
tivated.
113
Alarms inhibited.
164, 165
Sensor warm-up
phase has not
ended. Increased
measurement error
must be expected.
167, 170
Calibration interval
expired.
1)
Negative reading.
171
Value below range
minimum.
Low EC sensor
1)
163
vitality
®
1) Polytron
8100 EC only
Instructions for use
Remedy
Enter password again.
Remedy
Check supply voltage.
Have the instrument
checked by DrägerService.
Download the data as soon
as possible and clear the
data logger.
Reset the instrument to fac-
tory default settings. If this
error occurs again: Have
the instrument checked by
DrägerService.
Set date and time.
Have the instrument
checked by DrägerService.
Have the instrument
checked by DrägerService.
Deactivate SW dongle.
Enable alarms
Wait until the sensor has
warmed up. Do not calibrate
before sensor is fully
warmed-up.
Recalibrate the instrument.
Recalibrate zero point.
Replace sensor
®
|