Polytron 8200 Cat Drägersensor Dq; Polytron 8200 Cat Drägersensor Lc; Polytron ® 8310 Ir; Polytron ® 8700 Ir - Dräger Polytron 8000 Serie Notice D'utilisation

Masquer les pouces Voir aussi pour Polytron 8000 Serie:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

en
|
Factory default settings
Menu
Default setting
Hysteresis for A1
0
alarm at direction
falling
Hysteresis for A2
0
alarm at direction
falling
Calibration inter-
Depending on the
val [days]
sensor
Display capture
20.9 Vol% (O
offset
Display capture
-0.25 Vol% (O
low
Display capture
0.25 Vol% (O
high
Full scale deflec-
25 Vol% (O
tion
O
LS)
2
®
15.3.2 Polytron
8200 CAT DrägerSensor
Menu
Default setting
A1 Alarm
20 % LEL
A2 Alarm
40 % LEL
Hysteresis for A1
1 % LEL
alarm at direction
rising
Hysteresis for A2
1 % LEL
alarm at direction
rising
Calibration inter-
180
val [days]
Display capture
0 % LEL
offset
Display capture
-5 % LEL
low
Display capture
2 %LEL
high
Full scale deflec-
100 %LEL
tion
®
15.3.3 Polytron
8200 CAT DrägerSensor
Menu
Default setting
A1 Alarm
2.0 % LEL
A2 Alarm
4.0 % LEL
Hysteresis for A1
0.1 % LEL
alarm at direction
rising
Hysteresis for A2
0.1 % LEL
alarm at direction
rising
32
Range
0 to full scale
deflection -A1
0 to full scale
deflection -A2
0 to 720
)
0 to 25 Vol% (O
2
2
LS)
0 to 100 Vol% O
2
)
0 to -0.75 Vol%
2
)
0 to 0.75 Vol%
2
,
5 to 25 Vol% (O
2
2
LS)
5 to 100 Vol% O
2
®
DQ
Range
1 to 100 %LEL
0 to A1
0 to A2
0 to 360
-5 to 4 %LEL
-5 to 0 %LEL
0 to 4 %LEL
Range is not
adjustable
®
LC
Range
0.1 to 10 % LEL
0 to A1
0 to A2
Menu
Full scale deflec-
tion
Calibration inter-
val [days]
Display capture
offset
Display capture
low
Display capture
high
)
1) For firmware versions < 3.0.0
®
15.3.4 Polytron
Menu
)
A1 Alarm
A2 Alarm
Hysteresis for A1
alarm at direction
rising
Hysteresis for A2
alarm at direction
rising
Calibration inter-
val [days]
Display capture
offset
Display capture
low
Display capture
high
Full scale deflec-
tion
®
15.3.5 Polytron
Type 334
Menu
Default set-
ting
Calibration
Methane
gas
Calibration
%LEL
gas units
Calibration
50 %LEL
gas concen-
trations
A1 Alarm
20 %LEL
A2 Alarm
40 %LEL
Instructions for use
Default setting
Range
10 % LEL
5 to 10
1 to 10 % LEL
180
0 to 360
0 % LEL
-0.5 to 0.4 % LEL
-0.5 % LEL
-0.5 to 0 % LEL
0.2 %LEL
0 to 0.4 % LEL
8310 IR
Default setting
Range
20 % LEL
1 to 100 % LEL
40 % LEL
1 % LEL
0 to A1
1 % LEL
0 to A2
180
0 to 360
0 % LEL
-3 to 4 % LEL
-3 % LEL
-3 to 0% LEL
2 %LEL
0 to 4% LEL
100 %LEL
Range is not
adjustable
8700 IR
Range
0.01 to 100 Vol% at unit
%LEL = 0.3 to 100 %LEL
®
|
Dräger Polytron
1)
% LEL
8000 Series

Publicité

Table des Matières
loading

Table des Matières