Tecno-gaz KYRI DSS Mode D'emploi page 18

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

GB
The KYRI DSS suction equipment does not need maintenance or lubrication.
It is, however, necessary to inspect the unit before each use. With regard to training, given the information contained in the user
manual and since it is easy to understand the said device, it doesn't appear to be necessary.
The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or storage.
Always check the integrity of the footswitch power cord. Connect cable to electrical network and turn switch on.
Close the aspiration outlet with your finger and with suction regulator in maximum vacuum position check that the vacuum
indicators reaches -90 kPa (-0.90 bar) maximum. Rotate the knob from right to left and check the aspiration regulating control.
The vacuum indicator should go down -40 kPa (-0.40 bar). Verify that loud noises are not present, these can indicate wrong
functioning. A protection fuses (F 1 x 1.6 A L 250V) reachable from exterior and situated in the plug protects the instrument.
For fuses replacing, always the type and the range. Before changing the fuse, disconnect the plug from the power supply socket.
Internally, the device (only for devices fitted with a circuit board) is protected by a fuse (F 500mA L 250V) that cannot be reached
from the outside, so please contact a technician authorised by the manufacturer for its replacement.
Fault type
1.
No aspiration
2. No aspiration
3. The aspirator didn't
work
4. The Vacuum power on
the patient side is either
very low or absent
Faults 1 - 2 - 3 - 4
The hydrophobic, anti-backflow and antibacterial filter, positioned inside the disposable bag, protects the appliance from the passage
of any liquids, as well as acting as an overflow valve by deactivating the generation of the vacuum whenever the envisioned
maximum filling level is reached.
The hermetic sealing of the lid makes disposal particularly hygienic, easy and safe, both for the patient and the staff assigned to
removal or replacement of the component itself.
CA-MI Srl will provide upon request electric diagrams, components list, descriptions, setting instructions and any other information that
can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT
TECHNICAL SERVICE. THE MANUTACTURER DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE
TECHNICAL SERVICE CHECKING, APPEARS TO BE TAMPERED.
MAINTENANCE
Cause
System not fined suitably
System not fined suitably
Cable is damaged
External power source failure
a) Vacuum regulator set to
minimum
b) Connection tubes blocked,
kinked or disconnected
c) Shut-off valve blocked or
damaged
d) Pump motor damaged
None of the remedies has
achieved the desired results
Insert a bag and apply the lid to the re-usable rigid container,
pressing all around the perimeter, making sure the system is
sealed completely.
Apply the reusable conical "Male" connection.
Replace the cable
Check the external power source
a) Turn the vacuum regulator clockwise and check the value of
the vacuum on the gauge
b) Replace or reconnect the tubes, check the jar connections
c) Replace the Flovac 2000ml Liner Bag with a new one
d) Refer to authorised service personnel
Contact the seller or After-sales Assistance Service
18
Solution

Publicité

Table des Matières
loading

Table des Matières