This message appears in different circumstances - please find below the most common:
Causes of a Lo Message
Reading hampered by hair, sweat...
Reading hampered by an air flow...
The reading was taken too close, the ThermoFlash®
didn't have enough time to reset.
Measuring distance is too far.
26
Visiomed - LX-361T - USER MANUAL - 012017
Solutions
Make sure there is no obstruction.
Make sure you are not in a draught; this may interfere
with the infrared system.
Respect a delay of 15 seconds between each reading
- 1 minute is advised.
Respect a measuring distance between 2 to 5 cm,
away from the forehead.