VI. STARTING UP YOUR THERMOFLASH® LX-361T
- Insert the batteries.
- At the first use or after changing the batteries, wait around 10 minutes for the thermometer to warm up.
A
- Aim at the forehead (SEE SCHEMA
PAGE 2) at 2-5 cm (0.8-2 in), press the measure button and the tem-
pe-rature will be displayed immediately. The temperature can also be taken behind the ear on the jugular.
- Before taking a reading, make sure there is no hair or sweat on the forehead.
- Change the batteries when
shows on the screen.
VII. CONFIGURATION & MENU FUNCTIONS

1. CHOOSING HOW TO TAKE THE TEMPERATURE

The THERMOFLASH® LX-361T is specially designed to take a human body
temperature with the BODY function.
language
You can also use the ThermoFlash® to measure the temperature of the sur-
face of an object, a food or a liquid by selecting the SURFACE TEMP mode.
By selecting the ROOM mode, it will display the ambient temperature of
the room you are in.
MODE
°C/°F
MEM
Temperature range for BODY mode: 32°C - 42.9°C (89.6°F - 109.2°F).
Temperature range for SURFACE TEMP mode: 0°C - 60°C (32°F - 140°F).
Temperature range for ROOM mode: 0°C - 50°C (32°F - 122°F).
Press the Measure button then the MODE button for one second, the screen will display BODY.
Press the MODE button again to switch to SURFACE TEMP or ROOM mode.
Note: The THERMOFLASH® LX-361T is pre-calibrated to BODY.
Important: Surface temperature is not the same as internal body temperature. Internal temperature
must always be read with the BODY function.
Always select the BODY function for an internal temperature reading and the SURFACE TEMP or ROOM
function for an external reading (e.g. bottle, bath, bedroom...).

2. CHOOSING THE TEMPERATURE UNIT

Press the °C/°F button for one second: the THERMOFLASH® LX-361T will display the temperatures in
degrees Celsius (the screen will display °C).
Press the °C/°F again, and the THERMOFLASH® LX-361T will display the temperatures in degrees Fah-
renheit (the screen will display °F).

3. THE MEM BUTTON

F1. High temperature alert setting

Press and hold the MEM button for 3 seconds, and the thermometer will beep and the screen will show:
F1. The alert setting will flash on and off.
Press the °C/°F button to lower the setting by 0.1°C, or the button MODE button to increase it by 0.1°C.
To save the setting, press the MEM button.
Note: the high temperature alert is by default set to 38°C.

F2. Recalibration

To recalibrate the temperature readings of your THERMOFLASH® LX-361T, press and hold the MEM button
22
Visiomed - LX-361T - USER MANUAL - 012017
for 3 seconds. The thermometer will beep and display: F1. Press the MEM button again to display the F2 menu.
Press the °C/°F button to lower the setting by 0.1°C, or the MODE button to increase it by 0.1°C.
To save the setting, press the MEM button.
If you are unsure, we recommend that you leave the setting at +0.0°C (default setting).
Any time you move the ThermoFlash® to a place with a significantly different ambient temperature,
leave it at this temperature for at least 15 minutes before use.

4. EXIT CONFIGURATION MODE

Press and hold the MEM button until the screen switches itself off.

5. THE BACKLIGHT BUTTON

Press button 4 to activate the LED and to help you while taking a temperature at night.

6. THE LANGUAGE BUTTON

Use the language button to select the language for the voice function (French/English/Spanish/OFF).
The name of the language selected will be spoken. By default, the THERMOFLASH® LX-361T is set to French.

7. DATA SAVING

To display the last temperature readings saved, press the MEM button for 5 to 7 seconds.
You will then see the last temperature reading. Scroll through the readings with the MODE button.
The number displayed in the interval between two readings is the reading number.
Press the °C/°F button to scroll back.
To exit the memory menu, press the measure button. The ThermoFlash® will switch off automatically
after five seconds.

8. CHANGING THE BATTERIES

Display: When the LCD screen displays BATTERY
Operation: Open the cover and replace the old batteries, making sure to insert them the correct way
round. Inserting them incorrectly may damage the thermometer and void the thermometer's warranty.
Never use rechargeable batteries. Use single use batteries.
VIII. PRECISION & TECHNICAL CHARACTERISTICS
1. Normal conditions of use operating temperature: 10°C- 40°C (50° - 104°F). Normal conditions of sto-
rage and transportation temperature: -20°C - 50°C (-68° - 122°F). Humidity rate: 10 - 85%
2. Power: DC3V (2 batteries LR03).
3. Size with base: 172,5 x 60,35 x 60,35 mm (L x W x H).
4. Weight: 120g.
5. Display Resolution: 0.1°C (0.1°F).
6. Measuring range:
In BODY mode: 32°C - 42.9°C (89.6°F – 109.2°F).
In SURFACE TEMP mode: 0°C - 60°C (32°F - 140°F) and ROOM mode: 0°C - 50°C (32°F - 122°F).
7. Energy consumption: ≤ 50mW.
8. Accuracy: 36°C - 39°C : ± 0.2°C.
9. Measurement distance: 2 cm – 5 cm (0.8 in - 2 in).
10. Automatic standby: 5 sec.
This product complies with the electromagnetic compatibility requirements EN60601-1-2 and the safety standards
60601-1, as specified in the EEC directive 93/42/EEC.