Pump Set Dislodged Error (High Priority Alarm)
The PUMP SET DISLODGED screen will appear if the black ring retainer (MISTIC) is not properly
loaded in the MISTIC pocket in the pump set loading area. Check and correct the positioning of the
MISTIC retainer if possible. If the error cannot be resolved, load a new pump set, prime it, and press
CONTINUE to restart the feeding, Figure 30.
If the error cannot be resolved, press
pump into service.
Battery Low (Medium Priority Alarm)
Figure 31. Battery Low error
screen. Indicates that the battery
must be recharged immediately.
Feeding Complete (Information)
The FEEDING COMPLETE information screen appears after completion of the programmed
feeding, Figure 32. Press
to the SET LOADED opening menu, Figures 7A-7D.
Figure 33. Running screen with the >
24 HRS of Pump Set Usage Indicator in
upper left of the screen.

LED Array

The pump status LED array on the upper right of the pump gives a quick visual indication of the pump status, especially
in darkened rooms.
A green light status indicates normal pump operation.
A yellow light status indicates an informational or warning situation for a low battery condition, the pump in HOLDING mode, or
one of the following medium priority alarms: LOW BATTERY ERROR, HOLD ERROR, FEED ERROR, FLOW ERROR, and FLUSH
ERROR.
A red light status indicates a high priority alarm. The screen will show one of the following error messages: SYSTEM ERROR, PUMP
SET DISLODGED, or ROTOR ERROR.
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
POWER to stop operation of the pump and put a different
The BATTERY LOW screen appears and the alarm beeps continuously when the battery needs to
be recharged. There is approximately 30 minutes of battery life remaining when this screen
appears, Figure 31.
Plug the A/C power adapter in to begin recharging. The pump will automatically return to
the screen that was active prior to the error. The battery will charge continuously whenever
the pump is plugged into a wall outlet. 6 hours of charging is required to fully recharge the
battery pack.
If this screen appears while the A/C adaptor is plugged in to the pump, check to make sure the
A/C adaptor plug is pushed all the way in so it is flush with the back of the pump.
POWER DOWN to turn off the pump. Press
Set Use >24 Hours Warning
The pump set usage warning indicator will blink on the
RUNNING screen if a pump set has been used for 24 or more
hours (hours actually running). It is recommended to replace
pump sets after this length of usage. This icon is only an informational message and does not
require action.
Figure 33 shows an example of the RUNNING screen, where a feed & flush set has been in use
for more than 24 hours. The informational warning indicator shows up in the upper left on
the screen.
CONTINUE to return
29
Figure 30. Pump Set Dislodged
Error screen.
Figure 32. Feeding Complete
notification screen.