Continuous Mode (Adjust Feed); Continuous Mode Feeding; Continuous Mode Flushing - Covidien Kangaroo Joey Enteral Feed and Flush Pump with Pole Clamp Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 49
Continuous Mode — Adjust Feed

Continuous Mode Feeding

For CONTINUOUS feeding mode, the ADJUST FEED option is available in the opening menus (Figure 7A for feed-only sets or
Figure 7B for feed & flush sets). Select this option to program the feeding parameters from the ADJUST FEED screen, Figure 11A.
Press
FEED RATE in Figure 11A to define the rate of delivery. Press the
in increments of 1 mL/hr. Press
Press
FEED VTBD in Figure 11A to define the Volume To Be Delivered. Press the
from 1 to 3000 mL in increments of 1 mL. Press
will run until the supply is exhausted.
NOTE: When a FEED VTBD is set, the remaining volume to be delivered (remaining VTBD) will be displayed next to the
the lower right-hand portion of the screen (Figure 11A). If the FEED VTBD is reprogrammed during a feeding, from HOLDING mode,
the pump will restart the feeding and deliver the entire volume of the new VTBD setting.
VTBD symbol
Figure 11A. ADJUST FEED menu.
Continuous Mode Flushing (Not Available in EZMODE)
For CONTINUOUS feeding mode with a feed & flush set loaded, the opening menu will appear as shown in Figure 7B. Press
ADJUST FLUSH in Figure 7B to program the flushing parameters from the ADJUST FLUSH screen, shown in Figure 12A.
FLUSH VOLUME in Figure 12A to define the volume per flush occurrence. Press the
Press
the flush volume from 10 to 500 in increments of 1 mL. Press
FLUSH INTERVAL in Figure 12A to define the time interval between starts of flushing. Press the
Press
to program the time interval from 1 to 24 hours in increments of 1 hour. Press
The pump will automatically limit flushing settings which exceed pump capabilities to deliver.
Caution: The Kangaroo™ Joey enteral feed and flush pump with pole clamp flushes at a rate of 1960 mL/hr (32.7 mL/minute). Use
care when programming the flush volume so that it matches the patient's need.
Note: The pump will automatically flush an additional 25 mL of water after completing the programmed delivery of formula, as
long as at least one flush had occurred during formula delivery.
Figure 12A. ADJUST FLUSH menu,
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
ENTER (top button) to exit the menu.
ENTER (top button) to exit the menu. If the FEED VTBD is set to zero, the pump
Figure 12B. Setting the
continuous mode.
buttons in Figure 11B to program the rate from 1 to 400
Figure 11B. Setting the Feed Rate.
ENTER (top button) to exit the menu.
ENTER (top button) to exit the menu.
Flush Volume.
14
buttons in Figure 11C to program the volume
Figure 11C. Setting the Feed Volume
to be Delivered.
buttons in Figure 12B to program
buttons in Figure 12C
Figure 12C. Setting the Flush
Time Interval.
v
icon on
Table of Contents

Publicité

Table des Matières
loading

Produits Connexes pour Covidien Kangaroo Joey Enteral Feed and Flush Pump with Pole Clamp

Ce manuel est également adapté pour:

383400

Table des Matières