Selecting EZ Pump Mode

EZ PUMP MODE may be selected through the Biotech Options button which may be accessed on
the Biotech screen (see page 20). The EZ PUMP MODE setting will be stored in the machine pump
memory and saved when the unit is shut off.
Note: If the user inserts a feed/flush set while the EZ PUMP MODE is enabled, EZ PUMP
MODE will be cancelled and the unit will immediately revert back to the normal mode. If EZ
PUMP MODE is desired, the user must then re-enable EZ PUMP MODE through the Biotech
Options screen.
When enabled and not in CERTIFICATION MODE, a small EZ icon will be visible on the bottom
of the screen near the battery icon. However, it will not appear in the POWER DOWN, SYSTEM
ERROR, or LOW BATTERY SCREENS.

Selecting Continuous or Intermittent Feed Mode

The Kangaroo™ Joey enteral feed and flush pump with pole clamp can be programmed to feed continuously or intermittently.
The continuous feeding mode will deliver the enteral nutrition at a steady rate, either until the programmed volume has
been delivered or until the supply has been exhausted. The intermittent feeding mode delivers boluses of enteral nutrition at
programmed time intervals. The bolus volume and feed rate are also programmed.
To select continuous feed mode or intermittent feed mode, press
7A-7D, to access the MORE OPTIONS menu, Figure 10A.
The MORE option is available either before a pump set has been loaded (Figures 5A-5B) or after the pump set is loaded (Figures
7A-7D). See subsection "More Options" for information on the other options in the MORE OPTIONS menu.
CONTINUOUS/INTERMITTENT under the MORE OPTIONS menu, Figure 10A. Then press
Select
INTERMITTENT MODE from the SELECT MODE menu, Figure 10B. Press
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
Figure 10A. MORE OPTIONS menu.
MORE on the opening menu, shown in Figures 5A-5B or Figures
DONE to exit.
Figure 10B. SELECT MODE menu
for continuous or intermittent
mode selection.
13
Figure 9.
EZ Mode icon
CONTINUOUS MODE or