Figure 25. Hold Error screen.
Rotor Error (High Priority Alarm)
The ROTOR ERROR screen appears during RUNNING or PRIMING, when the pump has
detected that the pump set is no longer providing sufficient tension on the rotor. See
Figure 26. The ROTOR ERROR generally results from a problem with the pump set tubing
around the rotor.
Check that the pump set is not damaged, and re-load the tubing around the rotor as described
in subsection "Loading Pump Sets". Press
If the error cannot be resolved, press
different pump into service. Detection of this alarm condition may take up to 70 minutes to
occur at 1 mL/hr or up to 1 minute to occur at flow rates greater than 125 mL/hr.
Figure 27. Feed Error screen.
the feeding, Figure 27.
If the error still cannot be resolved, press
Detection of this alarm condition may take up to 360 minutes to occur at 1 mL/hr or up to 2 minutes to occur at flow rates greater
than 125 mL/hr.
Note: An occlusion may pressurize the feeding set, which can result in an unintended bolus of feed or flush solution when the
occlusion is cleared. This volume is less than 1ml.
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp
Hold Error (Medium Priority Alarm)
The HOLD ERROR screen, Figure 25, will appear if the pump has been inactive, without input, for
more than 10 minutes. See subsection "Hold" for description of HOLDING mode.
CONTINUE to return to the previous screen. Where settings can be adjusted, the pump
Press
can be set to run immediately, or the pump can be set to run in a specified number of minutes.
See Figures 18A-18B to see the options on the HOLDING menu screen.
If the error cannot be resolved, press
pump into service. Detection of this alarm condition may take up to 10 minutes to occur under
normal operating conditions.
CONTINUE to restart.
POWER to stop operation of the pump and put a
Feed Error (Medium Priority Alarm)
The FEED ERROR screen appears when the enteral formula is no longer being delivered because
the bag is empty or because of a clog between the pump and the bag. The pump determines this
by measuring the amount of fluid in the tube over time. Large amounts of foam or bubbles in the
feeding solution can also be a cause for this alarm. Check the bag to see if it is empty and re-fill
the bag as required. If the bag still contains feeding solution, remove cassette and check the bag
side tubing for excessive foam or bubbles. Clear bubbles from line and reload the feeding set or
replace with a new feeding set. Check the feed line to find the occlusion causing the blockage. If
the occlusion cannot be cleared, load a new pump set, prime it, and press
POWER to stop operation of the pump and put a different pump into service.
POWER to stop operation of the pump and put a different
27
Figure 26. Rotor Error screen.
Re-load the set tubing and
press CONTINUE.
CONTINUE to restart