BioTech Menu

The BioTech screen will show technical information about the pump, and has a lock feature to lock the pump settings against
unauthorized changes.

Accessing the BioTech Screen

To get the BioTech screen, press the power button to start the pump, and press and hold the top
sequence (kangaroo icon "hops" across the screen) executes. You must be pressing the top
order to access the BioTech menu. Figure 21 shows the BioTech screen.
Figure 21. BioTech screen.

BioTech Options

Figure 22. BioTech option screen,
with Lock Setting, EZ PUMP MODE,
and Communications features
activated. (Communication Mode
for service professionals only)

EZ PUMP MODE

EZ PUMP MODE to activate the EZ PUMP MODE feeding option. A check mark will appear next to the
Press
button to indicate that the feature is active. Pressing the
that the feature is off. If the EZ PUMP MODE feature is turned on, the EZ PUMP MODE icon will appear in the lower right corner of
the LCD next to the battery icon (see Figure 9). See page 13 for EZ PUMP MODE details.
Kangaroo™ Joey Enteral Feed and Flush Pump with Pole Clamp

BioTech Screen Data

SN is the serial number of the pump.
SW shows the software version number in this pump.
FLASH shows the version number of the flash data installed in this pump.
CERT INTV indicates the number of times the pump has been powered on since certification.
LOCK SETTINGS — Program Locked
To "lock" the settings against unauthorized changes, press
mark, as shown in Figure 22. The settings can only be changed again by pressing
SETTINGS to remove the check mark. If the settings are locked and a change is attempted, the
screen will indicate PROGRAM LOCKED, and will not allow access to the parameters, Figure 16.
The pump can be PRIMED or set to RUN, HOLD, etc. when settings are locked.
EZ PUMP MODE button again will remove the check mark indicating
20
button while the startup
button at the end of the sequence in
LOCK SETTINGS to get the check
LOCK
EZ PUMP MODE
Table of Contents