IKA HS 260 basic Mode D'emploi page 32

Masquer les pouces Voir aussi pour HS 260 basic:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

RMP_IN_X
X=4
RMP_IN_X_y
X=4
RMP_OUT_y n
X=4
hh:mm:ss
RMP_START_X
X=4
RMP_STOP_X
X=4
RMP_PAUSE_X
X=4
RMP_CONT_X
X=4
RMP_RESET_X
RMP_LOOP_SET_X
X=4
RMP_LOOP_RESET_X
X=4
IN_TYPE
IN_NAME
OUT_NAME
name
OUT_WD1@m
32
-87: not sufficient storage
space
Reading the real segment
number of ramp. With ramp
not started: 0
Reading the accumulated value
and the ramp segment duration
(hh;mm:ss) of ramp segment y.
Setting the accumulated value (n)
and the ramp segment duration
(hh:mm:ss) for ramp segment y
Starting the ramp function, begin-
ning with ramp segment No. 1.
(Only possible after prior
START_X.) After RMP_STOP_X
START_X is not necessary.
Switching off ramp function.
Rated value=0 (Ramp is main-
tained, that means , ramp can be
restarted with RMP_START_X.
Stopping the ramp function.
Freezing of real rated value and
real ramp segment time
Continuation of ramp function
(After prior RMP_PAUSE_X).
Switching off ramp functions
and deleting of all set ramp
segments
To work off the ramps in one
loop
Ending of ramp loop
Requiring identification of
laboratory instrument
Requiring the designation name
Output of designation name
(max.6 characters, default: IKA_S-)
Watchdog mode 1: If the WD1
result occurs, the agitation pro-
cess is turned off and Er2 is dis-
played. The watchdog time is set
to m (10...1800) sec with echo of
the watchdog time.
This command starts the watch-
dog function and must always be
sent within the time set by the
watchdog.
OUT_WD2@m
Watchdog mode 2: If the WD2
result occurs, the target speed
is set to the WD target safety
speed that has already been set.
The warning WD is displayed.
The WD2 result can be reset
with OUT_WD2@0.
The watchdog time is set to m
(10...18000) sec with echo of
the watchdog time. This com-
mand starts the watchdog func-
tion and must always be sent
within the time set by the
watchdog.
OUT_SP_42@n
Sets the WD target speed with
echo of the value set.
Communication between laboratory
instrument and PC
(HS/KS 260 control version only)
PC 5.1 Adapter IKA - Control
The Adaptor PC 5.1 splits the analog and serial signals. The analog
output signals are alotted to a 7-polar coupling according to the
Namur-recommendation; the serial signals to a 9-polar Sub-D-bushing
plug (RS 232 C).
WD

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Ks 260 basicHs 260 controlKs 260 control

Table des Matières