Replace a probe
To replace an electrode or temperature sensor, refer to
Figure 4 Junction box wiring
Probe
pH/ORP electrode
pH/ORP electrode
pH/ORP electrode
Temperature sensor
Storage
If the system is to be removed from operation for a period of time, keep the electrode in a storage
solution.
1. Fill the protective cap with 3 M KCl.
2. Install the protective cap over the electrode bulb.
Table 1 Wire descriptions
Wire
Black–rigid (center) wire
Green–copper internal shield
Green–copper external shield
Brown and white
Figure 4
and
Table
1.
Description
Measures pH or ORP
Reference
Protective earth ground
Pt100 sensor
English 11