Faults, cause, remedy
Fault
Flow test has failed.
Horn is too quiet.
Optical alarm signals not visible or poorly
visible.
Life signal is not functioning.
Gas detection instrument not detected.
X-am 5x00 switches off quickly; not
powered.
Device defect displayed.
62
Cause
Flow test has not been performed.
Holder (pump or X-am 5100) incorrectly
attached.
Volume is set too low.
Configuration or pattern incorrectly set.
Configuration is set incorrectly.
IR interface dirty.
Incompatible gas detection instrument.
Incorrect software version in the gas
detection instrument.
IR interface faulty, clip on gas detection
instrument not correctly positioned.
Holder is not correctly positioned.
Power contacts dirty or damp.
X-am 5x00 removed from X-zone during
operation.
Poor contacting with X-am 5x00.
Remedy
Perform flow test, repeat if necessary.
Reattach holder (pump or X-am 5100) and
ensure the correct position.
Set the volume using the Dräger CC-
Vision PC software (see Chapter 11 on
Page 60).
Configure the optical alarm signals using
the Dräger CC-Vision PC software (see
Chapter 11 on Page 60).
Configure the life signal using the Dräger
CC-Vision PC software (see Chapter 11
on Page 60).
Clean IR interface.
Use X-am 5x00.
Have the software update performed by
DrägerService.
Close the clip on the X-am 5x00.
Re-position holder and check that it is
correctly fitted.
Clean power contacts.
Acknowledge alarm on X-zone device,
switch off X-zone device.
Clean power contacts on the X-zone and
X-am 5x00.