12.1 Fault messages

Special symbol »
«
and displayed
numerical code:
01
02
03
04
05
06
07
08
09
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
29
30
31
32
Dräger X-zone 5500 / 5800
Cause
X-am 5x00 with alkali supply unit.
Interruption to communication with
X-zone device.
Communication error with
X-zone battery controller.
Main battery alarm X-am 5x00.
Battery pre-alarm X-am 5x00.
Charging current X-am 5x00 too low.
Holder (pump or X-am 5100) detected but no
pump installed.
Flow fault
Holder status change (pump or X-am 5100)
during running operation.
Check sum error program code
Check sum error operating parameter
Check sum error operating parameter
Check sum error operating parameter
Working memory test error
Faulty ADC conversion.
No contact with main device in grouping mode. Check wireless connection to group device.
Charging electronics faulty.
Battery completely discharged.
Battery main alarm of the X-zone device.
Battery pre-alarm of the X-zone device.
Device error X-am 5x00.
Alarm pattern X-am 5x00 faulty.
Incorrect software version in the gas detection
instrument.
Gas type or unit unknown.
More than 8 different gases in one group.
Modbus watchdog was not reset by the external
module.
Battery main alarm
Battery completely discharged.
Charging electronics faulty.
Communication error with X-zone battery
controller.
Remedy
Insert X-am 5x00 with battery power pack.
Check IR interface on X-zone and on X-am 5x00.
Contact DrägerService.
Check charging contacts on X-zone device and on
X-am 5x00.
Check charging contacts on X-zone device and on
X-am 5x00.
Check charging contacts on X-zone device and on
X-am 5x00.
Use holder for diffusion mode.
(Operation with X-am 5100 not possible.)
Check intake hose.
Holder pump: Check intake hose and pump filter.
Holder X-am 5100: Check gas inlet filter, cap, hose
and pump filter.
Check firm fit of the holder (pump or X-am 5100).
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Contact DrägerService.
Charge X-zone device.
Charge X-zone device.
Charge X-zone device.
Test X-am 5x00.
Contact DrägerService.
Have the software update performed by
DrägerService.
Test gas measuring devices individually to ensure
that all gases are known. Do not use gas
measuring devices with unknown gases.
Test sensor fitting within the group and replace gas
measuring devices if necessary or remove from the
group.
Check connection to external module or switch off
X-zone device.
Charge X-zone device.
Charge X-zone device.
Contact DrägerService.
Contact DrägerService.
Faults, cause, remedy