Carrying out a functional test with gas

WARNING
!
Risk to health! Never inhale the test gas.
Observe the hazard instructions of the appropriate
Safety Sheets.
CAUTION
!
A function test must be carried out before using any
device and then at regular intervals in accordance with
national and local guidelines.
Safe pump operation is only ensured when the
functional test is performed after each mounting of the
holder.
NOTICE
i
i
Sensor-specific features are described in the relevant
sensor data sheets.
The functional test can be carried out in 3 ways.
3.4.1
Functional test of the X-am 5000/5600
Perform a function test in accordance with the instructions
for use provided with the gas detection instrument before
using with the X-zone.
3.4.2
Function test of the X-am 5000/5600 in
combination with the X-zone
In diffusion mode
1. Switch on the X-zone (see "Switching on the instrument"
on Page 41).
2. Attach adaptor (1) (order no. 83 23 314) to the holder
(diffusion).
3. Connect the test gas cylinder hose to the adaptor (2).
4. Open the test gas cylinder valve to let test gas flow over the
sensors.
5. Wait until the device displays the test gas concentration
with sufficient tolerance:
Ex: ±20% of the test gas concentration
1
Upon application of the Dräger mixed gas (order no. 68 11 130)
the displays should be within this range. Different concentrations
can be set using the Dräger CC-Vision PC software provided.
Dräger X-zone 5500 / 5800
6. Close the test gas cylinder valve and remove the adaptor
If the displays are outside of the above-mentioned ranges:
Have the X-am 5x00 calibrated by the service personnel.
In pump operation (X-zone 5800 only)
1. Switch on the X-zone (see "Switching on the instrument"
2. Screw the valve on to the test gas cylinder.
3. Screw the hose from the test gas cylinder clockwise onto
4. Wait until the device displays the test gas concentration
1
2
5. Unscrew the hose on the test gas cylinder anti-clockwise
If the displays are outside of the above-mentioned ranges:
Have the X-am 5x00 calibrated by the service personnel.
3.4.3
!
02733112.eps
1. Switch on the X-zone 5800 (see "Switching on the
2. Hold the test gas source in front of the cap.
3. Open valve of test gas source to let test gas flow over the
4. Recommendation: Wait until the instrument displays the
1
5. Close valve of test gas source.
1
O
: ±0.8 vol.%
2
TOX: ±20% of the test gas concentration
Depending on the test gas concentration, alarm A1 or A2
is issued when the alarm thresholds are exceeded.
from the holder.
on Page 41).
the filter.
with sufficient tolerance:
Ex: ±20% of the test gas concentration
1
O
: ±0.8 vol.%
2
TOX: ±20% of the test gas concentration
Depending on the test gas concentration, alarm A1 or A2
is issued when the alarm thresholds are exceeded.
from the filter.
Function test of the X-am 5100 in combination with
the X-zone 5800
CAUTION
The gas inlet opening of the X-am 5100 and the cap
must not be covered or dirty to ensure a correct test.
instrument" on Page 41).
sensor.
test gas concentration with sufficient tolerance.
Wait at least until the alarm threshold A1 or A2 is
exceeded.
If the alarm thresholds are exceeded, the instrument
displays the gas concentration in alternation with » A1 « or
» A2 « depending on the test gas concentration.
Operation
1
.
03633112.eps
1
1
.
43