10. Establish plug connection to the bottom plate.
11. Position upper part of the housing onto the lower part
(note preferred position).
12. Tighten the screws (M5 cylinder-head screw with internal
hexagon) at the housing bottom (120 Ncm ±20 Ncm).
WARNING
!
Explosion hazard!
Do not dispose of used batteries into fire and do not
open by force.
Dispose of the batteries following the applicable
national regulations.
NOTICE
i
i
It is recommended to fully charge the device after the
battery block has been replaced.

13.4 Charging the batteries

WARNING
!
Explosion hazard!
Do not charge underground or in explosion-hazard
areas. The chargers are not designed in accordance
with the guidelines for firedamp and explosion
protection.
CAUTION
!
Persons wearing implants may only handle the
powered induction charger if their pacemakers and
active implants conform to the relevant legal
requirements. Dräger only guarantees conformity with
Directive 2004/108/EC.
NOTICE
i
i
The transmitter coil of the induction charger generates
a weak magnetic alternating field. During operation, all
requirements of the relevant standards regarding
electromagnetic faults are observed. The legal
requirements of Directive 2004/108/EC are met.
The warranty on the battery becomes null and void if
the device is not fully charged at least every 2 months
when not in use.
Dräger recommends storing devices that are not in use
in the induction charger (order no. 83 20 626).
To maintain the lifetime of the batteries, charging is
temperature controlled and only performed in a temperature
range of 5 to 35 °C. When this temperature range is left, the
charging is automatically interrupted and automatically
continued after the temperature range has been reached
again.
During the charging, the battery LED flashes red, red/green or
green at a frequency of one Hz, depending on the battery
status. As soon as charging is completed, the device changes
to maintenance charging. In this status, the battery LED
continues to flash green.
Dräger X-zone 5500 / 5800
13.4.1 Inductive charging
CAUTION
!
Fire danger/device damage!
Use only Dräger inductive chargers!
Do not place any metallic objects on the charger.
1. Connect the induction charger to the mains using the
device cable.
The control lamp on the charger is lit in green.
2. Place the device onto the induction charger.
The typical charging time is: <14 h - 24 Ah;
NOTICE
i
i
For good power transmission, it is not permitted for
there to be any contamination between the charging
cradle and the X-zone device.
13.4.2 Cabled charging
CAUTION
!
Fire danger/device damage!
Use only Dräger inductive charging accessories!
1. Plug the charging cable into the charging port on the rear
side of the device.
2. Connect the power pack to the mains.
The typical charging time is: <14 h - 24 Ah
NOTICE
i
i
Dräger recommends storing devices that are not in use
in the induction charger (order no. 83 20 626) or on the
plug charger (order no. 83 20 749).