General Notes; System Signals (Audible); Recovery; Switchover Time / Distance - Kettler SDIR Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

General notes

Sounds emitted by the system
Switch on
After switching on, a brief acoustic signal is given during the
segment test.
End of program
An acoustic signal is given at the end of a program (profile
programs, countdown).
Maximum pulse exceeded
If the set maximum pulse is exceeded by one beat, two brief
beeps are emitted every 5 seconds.
Error signal
In case of error, e.g. recovery not possible without pulse signal,
3 short beeps are emitted.
Automatic scan – switch function on and off
When the automatic scan function is activated and de-activated,
a short beep is emitted.

Recovery

During the function the flywheel mass is braked automatically.
Further exercising is not advisable.
Calculation of fitness mark (F):
Mark (F) = 6.0 –
P1 pulse during exercising,
F1.0 = very good,

Switchover time / distance

In programs 1- 6, the profile per column in the settings can
be switched from time mode (1 minute) to distance mode
(200 metres or 0.1 miles).

Profile display during exercising

At the start the first column flashes. During the sequence it moves
further to the right.
With profiles over 30 minutes up to the centre (column 15).
The profile then shifts from right to left. If the end of the profile
appears in the last column (column 30) the flashing column
moves further to the right until the end of the program is reached.
R.p.m.(speed)-related operation
(constant braking moment)
The electromagnetic eddy-current brake sets a constant braking
moment.

Average-value calculation

The calculation of the average takes place for each exercising
session.
Note on pulse measurement
The pulse calculation starts when the heart symbol in the display
flashes in time with your own pulse.
With chest strap
Refer to the appropriate instructions.
Malfunction of training computer
Press the SET button and keep pressed for some seconds.
(
)
2
10 x (P1–P2)
P1
P2 = recovery pulse
F6.0 = very poor
Note on interface
For interface possibilities, refer to our website www.kettler.net.

Glossary

Age
Input quantity for calculating the maximum pulse.
Control system
The electronics control the resistance or the pulse according to
manually input or pre-set values.
Dot display area
Display area with 30 x 16 dots for indicating power profiles,
text and values.
Energy (real.)
Calculates the energy expended by the body with an efficiency
of 16,7 % to provide the mechanical power. The remaining
83.3 % is converted by the body into heat.
Fat-combustion pulse
Calculated value of: 65 % MaxPulse
Fitness pulse
Calculated value of: 75 % MaxPulse
Glossary
A collection of attempted explanations
HI symbol
If "HI" appears, a target pulse is 11 beats too high. If HI flashes,
the Max-Pulse is exceeded.
Interface
9-pole SUB-D socket (RS232/serial) for exchange of data with a
computer.
LO symbol
If "LO" appears, a target pulse is 11 beats too low.
Manual
Calculated value of 40 - 90 % MaxPulse
MaxPulse
Calculated value of 220 minus age
Menu
Screen in which values can be input or selected.
Power
Current value of mechanical (brake) power in Watts which the
apparatus converts to heat.
Profile
The changes in resistance over time or distance as shown in the
dot display.
Programs
Exercising possibilities requiring resistances or target pulses to
be defined manually or automatically.
Pulse rate
Recording of number of heartbeats per minute

R.p.m.-related operation

The power indicated does not change until the number of strokes
per minute or the brake stage is changed. When the number
33

Publicité

Table des Matières
loading

Table des Matières