5.7.3

Peak value function

This function displays the highest load value (peak value) of a weighing.
Measuring frequency: 200ms
Attention:
Never exceed the maximum permitted load of the peak value on the
balance (!!Danger of breaking!!).
Switch on the balance, keep the HOLD-key pressed until the current setting „Hx"
(H1 – H6) appears.
Press ON/OFF button repeatedly until the setting „H6" is displayed.
Use the HOLD-key to confirm your selection
Suspend the material to be weighed
The peak value appears for a short time, symbolised by the [] top left. The
balance returns automatically to zero and is ready for further measurement.

6. Menu

When the balance is switched off, press the HOLD-key and keep it pressed
Do not release the HOLD-key. Press also the ON/OFF/TARE-key and keep it
pressed
Keep the ON/OFF/TARE-key pressed, however release the HOLD-key
Press HOLD-key anew
Keep both keys pressed until in the display "tr" appears
Release both keys. The balance is situated in the menu.
Using the ON/OFF/TARE button you can select between the following functions:
• tr (Zero tracking) : ON/OFF
• AF (Auto off): ON/OFF
• rST (reset to factory settings): YES / NO
Press the HOLD button to confirm the selected function
In the display appears the current setting „ON" or „OFF" resp. „YES" or „NO". You
can select between „ON" or „OFF" and/or „YES" or „NO" with the
ON/OFF/TARE button. Confirm your selection with the HOLD-key. After a short
period the balance returns automatically to weighing mode.