Controller parameters

The main menu for adjusting the Multronic can be accessed through operating the ↑-key
in the measuring mask.
Main menu
The controller parameters function can be selected through placing the cursor onto the
selected item and operating the ENTER key.
Controller parameter
module selection
(not available at
Multronic OC!)
The conductive C measuring module that is to be parametered can now be selected. To
do so, move the cursor onto the corresponding selected item and operate the ENTER
key.
The adjustment of the conductive C controller parameters is carried out on two pages.
The display of the first page depends on the controller type setting in the module
configuration menu.
The following settings apply to the signal device:
Conductive
C controller
parameters
page 1 (signal device)
The set point (W) function is used to set the nominal value within the measuring range.
Value range (depending on selected measuring range):
Measuring range Value range
0..2µS
0..20µS
0..200µS
The switch difference (XSD) function is used to specify the switch hysteresis around the
set point. The stated percentage refers to the nominal value.
Value range: 0.0 % to 30.0 %
417101802_Anhang02_LF_konduktiv.doc
Conductivity Measuring Module (contact measurement)
0.000 µS to 2.000 µS
0.00 µS to 20.00 µS
0.0 µS to 200.0 µS