It must be ensured that the selected measured value for the maximum output current is
larger than the measured value for the minimum output current. If the current measured
value falls below the measured value specified for the minimum output current, the
output current stays at this minimum. The same applies if the current measured value
rises above the allocation for the maximum output current.
Temperature compensation function can be adjusted either manually or automatically.
This can be selected in the first setting position of the temperature compensation
settings. If the auto function is selected, no additional temperature adjustments are
possible any more – the temperature that is to be set disappears. If the measured
value is to be compensated with a manually selected temperature, the setting man
must be selected and the temperature that is to be used for the compensation must be
entered in the second setting position (accessed with the →- key).
Value range: 0 °C to 99 °C (32 °F to 212 °F)
The module selection function can be used to return to the measurement parameter
module selection.
The value ranges for the voltages of a ± 500 mV redox measuring module must be
changed to - 500 mV to + 500 mV.
4.3

Controller parameters

The main menu for adjusting the Multronic can be accessed through operating the ↑-
key in the measuring mask.
Main menu
The controller parameters function can be selected through placing the cursor onto the
selected item and operating the ENTER key.
Controller parameter
module selection
(not available at
Multronic OC!)
The redox measuring module that is to be parametered can now be selected. To do so,
move the cursor onto the corresponding selected item and operate the ENTER key.
The adjustment of the redox controller parameters is carried out on two pages.
The display of the first page depends on the controller type setting in the module
configuration menu.
417101806_Anhang04_Redox.doc
Redox Measuring Module
- 33 -
Rev.: 06-07/10