The calibration start function is used to start the calibration process.
The go on function terminates the calibration process and opens the calibration module
selection menu.
Rinsing with buffer
solution
This is when the measuring probe must be immersed in the buffer solution at the latest.
Once this has been ensured the calibration can be continued through selecting
ENTER.
Measurement
This is where an automatic mV signal measurement of the reference buffer solution is
carried out.
The display of the measurement voltages makes it easier to estimate the measurement's
duration. If the measured value does not change anymore, it can be assumed that the
measuring process will be completed shortly and that the calibration will continue
automatically.
The calibration process can be cancelled through operating ENTER.
Calibration OK
If the calibration has been completed without any errors, this is indicated with the
calibration OK message. The calibration data is saved and the calibration module
selection menu is opened through operating ENTER.
Probe deviations too
high
The above warning is issued if a deviation of more than ± 60 mV from he reference
solution has been established during the measurement.
417101806_Anhang04_Redox.doc
Redox Measuring Module
- 39 -
Rev.: 06-07/10