The switching hysteresis around the limit contact can be specified in % with the limit-
contact (X2SD) function.
Value range: 0.0 % to 30.0 %
The controller of a different module can be specifically turned off with the limit controller
function as soon as the limit range has been reached.
Value range (module on plug-in connection 1): none / all / module 2 / module 3
Through selecting the module selection function one returns to the controller parameter
module selection.
The page 1 function is used to display the specific parameters of the selected controller.
4.4

Calibration

The main menu for adjusting the Multronic can be accessed through operating the ↑-key
in the measuring mask.
Main menu
The calibration function is selected through placing the cursor onto the selected item and
operating the ENTER key.
Calibration module
selection
(not available at
Multronic OC!)
The conductive C measuring module that is to be calibrated can now be selected. To do
so, move the cursor onto the corresponding selected item and operate the ENTER key.
Conductive C
calibration
417101802_Anhang02_LF_konduktiv.doc
Conductivity Measuring Module (contact measurement)
- 43 -
Rev.: 05-07/10