The measuring range dependent simulation resistance can now be put into the measuring
probe.
The value of the resistance depends on the measuring range:
Measuring range
0 ..
2 µS
0 .. 20 µS
0 ..200 µS
The calibration can now be continued through selecting ENTER.
Measurement 2
Measurement 2 takes place automatically and continues until the measurement range is
stable.
Calibration OK
Faultless completion of the calibration is indicated with the message 'calibration OK'. The
calibration data is saved through operating ENTER which also opens the calibration
module selection menu.
Unstable measured
value
This status signal appears if the measured value does not move within a limited range for
a period of time.
The calibration is cancelled through selecting the interrupt calibration function, the values
are not adopted and the calibration can be started anew.
The continue calibration function is used to complete calibration with the unstable values.
417101802_Anhang02_LF_konduktiv.doc
Conductivity Measuring Module (contact measurement)
Resistance
50.000 Ω
5.000 Ω
500 Ω
- 45 -
Rev.: 05-07/10