Defining Your Own Configuration Combinations - HBM MP30DP Manuel D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

PME-MP30DP

6.2.1 Defining your own configuration combinations

Only one configuration entry is available. The special identification format
(special format) must be used for this. The manufacturer-specific data
specifies the content and thus also the length of the input data and is 2 bytes
in length.
CFG entry no.
Meaning
0
channel 1
The following input and output data can be configured for the cyclical data
traffic. The choice of which data is actually transferred is communicated via
the manufacturer-specific data of the special identification format.
Configuring
manufacturer-spe
cific data
Byte no.
Bit no.
0
0
0
1
0
2
0
3
0
4
0
5
0
6
1
0
1
1
1
2
1
3
1
4
Tab 6.11: Selecting the data content via the manufacturer-specific data
The length of the input data is the sum of all the data lengths selected for the
communication in words. When selecting the 32 bit format and the float format
for measured values, the length values must be used in brackets.
A0737-4.0 de/en/fr
special format with inputs and outputs, maximum 9 words
output, maximum 13 words input, 2 bytes comment length
Length of cyclical data
inputs
(words)
1(2)
1(2)
1(2)
1(2)
1(2)
1
2
Contents
(data)
Length of cyclical data
outputs
(words)
1
1(2)
1(2)
1(2)
1(2)
33
Cyclical data content
Input values:
Gross
Net
Max
Min
Peak-to-peak
Status1
Status2
Output values:
Control word
Limit value level 1
Limit value level 2
Limit value level 3
Limit value level 4
HBM