Operating And Display Functions - Renfert Easyclean Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
4.6 Operating and display
functions
see also Figure 3
Action
switch on unit
switch off the unit
start ultrasound
- now -
start ultrasound
- temperature-controlled*);
with mixing of cleaning liquid
*) if set temperature > actual
temperature
stop ultrasound before end
of set period
switch on heating
switch off heating by hand
switch on sweep function*)
*) sweep and degas cannot be
operated at the same time
switch off sweep function
switch on degas function*)
*) sweep and degas cannot be
operated at the same time
switch off degas function
switch on auto degas func-
tion*)
*) sweep and degas cannot be
operated at the same time
EN
Setting
Result
• press on/off key
• unit is ready for
• press on/off key
• unit is switched off
• select period by turn-
• ultrasound is
ing knob for cleaning
period
• press key ►■
(ultra-sound)
• set period
• heating operates
• set temperature
• ultrasound is started
by turning knob for
temperature
• keep key ►■
pressed for > 2 sec
• set period ultrasound
• turn set period to 0
• ultrasound switched
or
• press key ►■
• select set
• heating operates
temperature
• turn set temperature
• heating switched off
to position „0"
• select set period
• ultrasound operates
• press key ►■
• press key sweep
• press key sweep
• sweep function is
• ultrasound continues
• select set period
• ultrasound operates
• press key ►■
• press key degas
• press key degas
• degas function is
• ultrasound continues
• press key ►■
• ultrasound operates
• keep degas key
pressed > 2 sec
Display
• on/off LED is on
operation
• all displays are off
• ultrasound LED is on
operating
• set period LED is on
• remaining period LED blinks
(only in timer operation)
• ultrasound LED blinks
• set period LED blinks
automatically after
• as soon as the set temperature is
reaching the set tem-
reached the ultrasound LED is on
perature
• set period LED is on
• remaining period LED blinks
runs down
• ultrasound LED is off
off
• set period LED is on
• set temperature LED is on
• actual temperature LED blinks and
goes towards set temp.
• as soon as actual temp. = set temp.,
only the set temp. LED is on
• if actual temperature > set tem-
perature, the temperature LED starts
blinking again
• set temperature LED blinks
• sweep LED is on
in sweep mode
• ultrasound LED is on
• set period LED is on
• remaining period LED blinks
• sweep LED is off
switched off
• ultrasound LED is on
• set period LED is on
in standard operating
• remaining period LED blinks
mode
• degas LED is on
in degas mode
• ultrasound LED is on
• set period LED is on
• remaining period LED blinks
• degas LED is off
switched off
• ultrasound LED is on
• set period LED is on
in standard operating
• remaining period LED blinks
mode
• degas LED blinks
in auto degas mode
• ultrasound LED blinks
for 10 minutes and
then switches off
- 4 -

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

1850-00001850-1000

Table des Matières