Example: 6 load cells require approx. 370 mA (this corresponds to the mini
mum current required by the power supply unit), the required
cable length is to be 100m. At 15VDC, this can be implemented
with a cable cross--section of 0.14mm
cross--section of at least 0.25mm
Power supply table
(when using the HBM interface converter set SC232/422B with an internal
consumption of approx. 70 mA):
Number
Number
Sum
Sum
of load
(
(current con-
cells
ll
sumption) *
ti
1
120 mA
2
170 mA
3
220 mA
4
270 mA
6
370 mA
8
470 mA
10
570 mA
12
670 mA
16
870 mA
* incl. HBM interface converter set SC232/422B
NOTE
The plug- -in power supply unit , included in the scope of delivery for the
HBM interface converter set SC232/422B supplies 15 VDC/530 mA. So it
can be used for operation with max. 8x C16i... load cells.

9.3 Command set C16i...

Please request the separate documentation of our AD104C electronics, if re-
quired. Operating instructions are available in German and English language.
10

Options

D Cable length 20 m (E
D Cable length 40 m (E
D 20 m metal braided cable, (E
A1461--1.2 de/en/fr
between converter and junction box
Nominal voltage 12 VDC
) *
Wire cross--section
2
0.14mm
0.25mm
352 m
500 m
176 m
314 m
117 m
210 m
88 m
157 m
59 m
105 m
44 m
79 m
35 m
63 m
29 m
52 m
22 m
39 m
= 20 t + 30 t)
max
= 20 t ... 60 t)
max
max
2
. At 12VDC, a minimum
2
needs to be selected.
Maximum cable length
Nominal voltage 15 VDC
2
2
0,5mm
0,14mm
500 m
500 m
500 m
296 m
419 m
197 m
314 m
148 m
210 m
100 m
157 m
74 m
126 m
59 m
105 m
49 m
79 m
37 m
= 20 t ... 60 t)
Wire cross--section
2
2
0,25mm
0,5mm
500 m
500 m
500 m
500 m
352 m
500 m
264 m
500 m
176 m
352 m
132 m
264 m
106 m
211 m
88 m
176 m
66 m
132 m