4.2
Cleaning and disinfecting
Clean and disinfect the full face mask after every use.
WARNING
!
The mask body must be cleaned sufficiently thoroughly that
no residues of skin care products remain on the mask body.
Otherwise the elasticity in the sealing area can be lost and the
full face mask will no longer sit correctly.
CAUTION
!
Do not use solvents (e.g. acetone, alcohol) or cleaning agents
containing abrasive particles for cleaning and disinfection.
Only clean the mask using the described methods and
cleaning/disinfecting agents. Other agents, dosages and
contact times can damage the product.
Do not clean or dry the inside of full face masks with visors
marked "PC/CC" with a cloth; this may damage the coating.
Full face masks with these visor types should be left
unpackaged after drying for at least a day in a normal climate
according to DIN EN ISO 291 so that the coating can
regenerate and the optimum service life can be achieved.
For information about suitable cleaning agents and
disinfectants and their specifications, refer to document
9100081 at www.draeger.com/IFU.
1. If necessary, remove accessories and clean them separately.
2. Prepare a cleaning solution consisting of water and a cleaning
agent.
3. Clean all parts using a soft cloth with cleaning solution.
4. Thoroughly rinse all parts under running water.
5. Prepare a disinfectant bath consisting of water and a disinfectant.
6. Put all parts to be disinfected in the disinfectant bath.
18
7. Thoroughly rinse all parts under running water.
8. Dry all parts in the air or in a drying cabinet (temperature: max.
60 °C). Do not expose to direct sunlight.
9. Mount the accessories, if necessary.
NOTICE
i
i
The full face masks with PC disc can also be cleaned and
disinfected in the washing machine. Further information can
be obtained from Dräger.
4.3
Checks
Check the full face mask according to the inspection and maintenance
intervals.
4.3.1
Visual inspection of the speech diaphragm
The speech diaphragm and O-ring must be clean and undamaged.
Otherwise clean or replace (see Chapter 4.4.4 on Page 21).
4.3.2
Visual inspection of the inhalation valve
The inhalation valve can be seen in the connector through the
visor. Check if both blades of the valve disc move freely by blowing
into the connector. Replace the valve disc, if required (see
Chapter 4.4.6 on Page 21).