Message
Out of range
Wrong password
Passwords do not match, try again
Program memory error
Data memory error
No matching data found in memory
Sensor ID already exists, previous SN
will be overwritten
9.1  E rror limits
Message
Out of range
44 Error messages
51710917_V02.2011_BA_SG9_98_EnFrEs.indb 42
Description and Resolution
Either entered value is out of range.
Enter a value, which is within the range shown on
the display.
or
Measured value out of range.
Make sure the electrode wetting cap has been re-
moved and that the electrode is properly connected
and placed in the sample solution.
If no electrode is connected, put the shorting clip in
the socket.
The entered PIN is not correct.
Re-enter the PIN.
Reset to factory settings, all data and settings will
be lost.
The confirmation PIN does not match with the entered
PIN.
Reenter PIN.
Meter recognizes internal error during start-up.
Switch the meter off and back on.
Contact METTLER TOLEDO service if problem per-
sists.
The data could not be stored in the memory.
Switch the meter off and back on.
Contact METTLER TOLEDO service if problem per-
sists.
The entered filter criterion does not exist.
Enter a new filter criterion.
Two sensors with the same ID but different SN are not al-
lowed in the meter. If a different SN has been entered for
this sensor ID previously, the old SN will be over w rit t en.
Enter a different Sensor ID in order to keep the pre v i-
ous ID and SN.
Range not accepted
pH
mV
[O
]
2
[O
]
2
[O
]
2
<-2.000 or > 19.999
<-1999.9 or > 1999.9
< 0.1% or > 600 %
< 0.01 mg/L or > 80 mg/L
< 0.01 ppm or > 80 ppm
21.04.2011 11:46:19