Connecting an IP67 sensor
To connect the IP67 sensor, make sure that the plugs are properly inserted. Twist the RCA
(Cinch) / mini LTW plug to ease the attachment of the sensor.
Connecting an ISM® sensor
ISM
sensor
®
When connecting an ISM
the calibration data to be transferred automatically from the chip of the sensor into the meter and is
used for further measurements. After attaching the ISM
the meter must be switched on.
If the meter is already switched on, the READ key is pressed.
If the meter is already switched on, the CAL key is pressed.
We strongly recommend you to switch off the meter when disconnecting an ISM sensor. In doing
so, you make sure that the sensor is not removed while the instrument is reading data from or writ-
ing data to the ISM-chip of the sensor.
The ISM icon
appears on the display.
The calibration history, the initial certificate and the maximum temperature can be reviewed and
print e d in the data memory.
3.3  O ptiOx™ BOD adapter and protective guard
Installing the BOD adapter or the protective guard
Unscrew and remove the thread ring from the InLab
Slide the BOD adapter or the protective guard over the front of the InLab
onto the sensor.
3.4  F itting the wrist strap
10 Installation
51710917_V02.2011_BA_SG9_98_EnFrEs.indb 8
sensor to the meter, one of the following conditions have to be met for
®
appears on the display and the sensor ID of the sensor chip is registered and
The special BOD (biochemical oxygen demand) adapter enables
quick and easy measurement in all current types of BOD bottles
with the InLab
OptiOx.
®
Thanks to the adapter the sensor reaches into the bottle only as
far as needed so that less water is being displaced during mea-
sure m ent. According to the EPA (Environmental Protection Agency,
USA), stirring is not necessary thanks to RDO technology.
The robust protective guarde made of stainless steel provides opti-
mum protection even in harsh environments.
Thanks to the additional weight of the protective guard it also acts
as a sinker taking the InLab
tions.
Fit the wrist strap as shown in the diagram.
sensor,
®
OptiOx to deeper measurement loca-
®
OptiOx. Keep it for later use.
®
OptiOx and screw it
®