7.3  C alibration data
Calibration data can be reviewed, transferred and deleted. Up to 5 calibrations per sensor ID are
stored in the memory.
Review
1
Select between the sensor types: pH, ion or DO (only SG98).
2
Press Review.
3
Select a sensor ID from the list and press Review.
4
Press
— or —
Press and hold CAL for 3 seconds in the single channel measurement screen.
5
Press Trans to send the displayed calibration data over the IR interface to a printer or PC.
Transfer
1
Select between the sensor types: pH, ion or DO (only SG98).
2
Press Trans.
3
Select a sensor ID from the list and press Trans.
4
Press
5
Press Trans to send all the calibration data of the selected sensor ID over the IR interface to a
print e r or PC.
Delete
1
Select between the sensor types: pH, ion or DO (only SG98).
2
Press Delete.
3
Select a sensor ID from the list and press Delete.
4
Press Yes when the message "This will delete all your selected data. Please confirm." appears
— or —
Press No to cancel and exit.
An active sensor ID cannot be deleted.
This menu is protected by a deletion PIN code. Upon delivery, the PIN code is set to 000000.
Change the PIN code to prevent unauthorized access.
51710917_V02.2011_BA_SG9_98_EnFrEs.indb 35
A list of calibrated sensor IDs appears.
and
to navigate between the previous or next calibration data sets.
The current calibration data is displayed.
A list of calibrated sensor IDs appears.
and
to navigate between the previous or next calibration data sets.
A list of sensor IDs appears.
After deletion, the sensor ID disappears from the list in the sensor ID menu.
Data management
37