10.6. Automatic controller (PLC) link cable
E
Extra flexible cable. Wire cross-section: 0.05 mm2 (26 x 0.05 mm Ø, AWG 30).
Wire colour
Black
Red
Blue
Green
Yellow
White
Brown
Pink
Grey
Braid
If more than one digital indicator is connected to the PLC via the bus lines, all GND, all VCC, all RS-485-B (D-), all RS-485-A (D+) and all PE wires must be tied together.
Signal name
GND (0 volt)
VCC
Output 3
Output 2
Output 1
Input 2
Input 1
RS-485-B (Data-)
RS-485-A (Data+)
PE
18
Signal description
Supply (0 volt)
Supply (9 ... 24 VDC)
Note: supply from a USB port (+5 V) is only
possible temporarily for manual or Proximity
cable configuration
Outside upper tolerance (Re-work)
Within tolerance limits (OK)
Outside lower tolerance (Scrap)
Favourite button function execution
Preset / Zero recall
Bus Link - B (half duplex)
Bus Link - A (half duplex)
Shielding (earth protection)