The display language is selected when the meter is powered on for the
first time.
1. Use the
or
to select a language from a list.
2. Confirm with
. The measurement screen shows DATA OUTPUT.
3. Select Deactivated if no printer or PC is connected and confirm.
Refer to
Select the data output
about Data Output.

Standard operation

Setup for dissolved oxygen
Before a calibration or a measurement for dissolved oxygen, the probe
must be polarized and the settings for atmospheric pressure and salinity
must be entered.
Polarize the electrode
If the probe or batteries are disconnected, connect the probe or install
the batteries and wait for polarization:
Disconnect time
< 5 minutes
5 to 15 minutes
> 15 minutes
Calibration
W A R N I N G
Chemical exposure hazard. Obey laboratory safety procedures and wear all of
the personal protective equipment appropriate to the chemicals that are handled.
Refer to the current material safety data sheets (MSDS) for safety protocols.
Calibration settings
The calibration settings contain Calibration type, Calibration frequency
and Display options.
10 English
on page 13 for more information
Polarization time
10 minutes
45 minutes
6 hours
1. From the main menu use the
Confirm.
2. Use the
to enter the calibration menu.
3. Use the
or
to select the following options:
Option
Description
Stability C.:
Criterion by stability—select Fast, Standard or Strict.
Calibration
pH: Calibration type—select Technical buffers,
type
DIN19266 Buffers, User Buffers, To a X value, data
introduction or Theoretical calibration.EC: Calibration type
—select Molar Standards, Demal Standards, NaCl
Standards, Calibration to a X value, data introduction or
Theoretical calibration. Refer to
information.DO: Calibration type—select Water saturated
air, Water saturated air & Zero, To a X value or Theoretical
calibration.
Cal. frequency
Calibration reminder—can be set between 0-23 hours (pH)
or 0–7 days (EC). The display shows the remaining time to
the new calibration. Refer to
on page 12 reminder for more information.
Display mV
Display mV—select YES or NO to show mV.
Display nA
Display nA—select YES or NO to show DO current in nA
(nanoampere).
Calibration types
Different calibration types can be selected.
1. From the main menu use the
Confirm.
2. Use the
to enter the calibration menu.
3. Use the
or
to select Calibration type.
Option
Description
Technical buffers
pH 2.00, 4.01, 7.00, 9.21 and 10.01 at 25 °C
(77 °F)
or
to select CALIBRATION.
Calibration types
for more
Set the calibration reminder
or
to select CALIBRATION.