S The PMX amplifier channel is set to the full bridge
sensor type, with the measuring range of 4 mV/V.
S Amplification is set to 1,000 N, for a sensor sensitivity
of 2 mV/V. If the sensor has TEDS sensor detection,
the amplifier channel can be parameterized automat­
ically. Activation takes place on the 2nd page of the
applicable amplifier settings.
S The filter type is Bessel, the filter frequency is set to 5
Hz (average to high attenuation).
S The data are now changed in the PMX and are dis­
played by a floppy disk symbol in the status line .
S Press this symbol to store the settings power failsafe
in the PMX (security prompt).
Example: Configuring the PMX with strain transducer
Example: SLB700A
Slot 2 is fitted with a PX455, its third channel is assigned
to the strain transducer.
A3259-4.0 HBM: public
Quick start