Caution: Do not pull distal end of catheter through incision as contamination of
wound may occur.
10.
Apply pressure to proximal tunnel for approximately 10-15 minutes or until
bleeding stops.
11.
Suture incision and apply dressing in a manner to promote optimal healing.
12.
Check catheter integrity for tears and measure catheter when removed. It
must be equal to the length of catheter when it was inserted.
References:
1. Zaleski GX, Funaki B, Lorenz JM, Garofalo RS, Moscatel MA, Rosenblum
JD, Leef JA. Experience with tunneled femoral hemodialysis catheters. Am
J Roentgenol. 1999 Feb; 172(2):493-6.
Medcomp
WARRANTS THAT THIS PRODUCT WAS MANUFACTURED ACCORDING TO
®
APPLICABLE STANDARDS AND SPECIFICATIONS. PATIENT CONDITION, CLINICAL
TREATMENT, AND PRODUCT MAINTENANCE MAY EFFECT THE PERFORMANCE OF
THIS PRODUCT. USE OF THIS PRODUCT SHOULD BE IN ACCORDANCE WITH THE
INSTRUCTIONS PROVIDED AND AS DIRECTED BY THE PRESCRIBING PHYSICIAN.
Because of continuing product improvement, prices, specifications, and model availability
are subject to change without notice. Medcomp
or contents without notice.
Medcomp
, Vascu-Sheath
®
Components, Inc.
14F x 28cm PRESSURE
ml/MIN
VENOUS
38 mmHg
ARTERIAL
-30 mmHg
16F x 28cm PRESSURE
VENOUS
26 mmHg
ARTERIAL
-20 mmHg
FLOW RATE TESTING REPRESENTS OPTIMUM
LABORATORY CONDITIONS.
, and Split Cath® are registered trademarks of Medical
®
200
300
ml/MIN
65 mmHg
-56 mmHg
200
300
ml/MIN
ml/MIN
49 mmHg
-43 mmHg
WARRANTY
reserves the right to modify its products
®
-8-
400
ml/MIN
103 mmHg
-90 mmHg
400
ml/MIN
78.7 mmHg
-72.4 mmHg