• When the chair is moving upwards or downwards or tilting do not place feet on the
sub-frame; this will prevent any jamming issues.
• Close the armrests with front closure before moving the Reflex upwards or tilting it.
• Make sure that the client always places his / her arms on the appropriate armrest in
order to avoid anything becoming jammed.
• Keep the armrests or front closure down / closed when moving the Reflex.
• Be aware of obstacles when moving upwards and downwards or tilting.
• Only relocate and move the Reflex in the lower position.
• When moving the Reflex, do not adjust upwards or downwards or tilt.
• Always try to move backwards over obstacles (rather than forwards).
• Set the Reflex to the correct height before using with a toilet.
• It is not permitted to charge the Reflex with more than 135 kg / 300 lbs.
• The Reflex may not be used to transport loads other than people.
• Only use original parts, supplied by Lopital.
• The use of components (chargers, cables, actuators, batteries, etc.) that have not been
supplied by Lopital may result in increased emissions or decreased immunity of the
Reflex.
• Use the Reflex in a clean and tidy working environment.
• Only verifiably authorised personnel is allowed to work with the Reflex.
• Charge battery in a well-ventilated, dry space.
• Swimming pool locations are not considered to be 'normal conditions' and will sub-
stantially shorten the lifespan of certain components.
• The Reflex cannot be used on a surface with a gradient over 5 degrees laterally.
• The Reflex cannot be used on a surface with a gradient of over 10 degrees forwards /
backwards.
• During use of the Reflex, the water temperature may not exceed 43°C / 109.4°F.
• The Reflex may only be used indoors.
• Portable and mobile RF communication equipment can affect medical devices.
• Medical electrical devices involve special EMC precautions and must be used as per
the EMC information outlined in the instructions for use.
• T he chair must not be used adjacent to or stacked on other products. However, should
that be necessitated, it must be checked whether the chair functions normally in this
configuration too.
Maintenance
• If the charger or cable become damaged, they must be replaced.
• The Reflex must be inspected and serviced at least once a year.