Any malfunctions during operation will be identifi ed by an error message on the display (D), see fi g. 5.
Proceed as follows in such cases:
Switch off device using the main switch (B)
Carry out corrective measures
Restart device
Error message Cause
E 2
E 3
E 9
E 21
E 23
E 24
E 25
E 26
E 28
E 29
E 30
E 32
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA
- send the device for repair, including a short description of the fault.
HB 10.1 Spray guard
Heruntergeladen von
manualslib.de
- In remote operation (PC), no communication
between RV 10 and HB 10
- IR connection to RV 10 disconnected
- Temperature inside the device is too high
- Fault in storing the safety circuit set
temperatures
- Memory chip (EPROM) defective
- Safety relay does not open
- Adjustable safety circuit defective
- Set safety temperature exceeded
- The heater control circuit switch (TRIAC) is
faulty. The heater or the supply line is
disconnected.
- Dry running
- Breakage of the controller sensor
- Short-circuit in the safety sensor
- Short-circuit in the controller sensor
- Error in calibration
®
service department,
Handbücher-Suchmachiene
Error message
Effect
Correction
Heating off
- Switch off the device
- Clean the IR interface or
- Switch on the device
Heating off
- Switch off the device
- Let the device cool down
- Switch on the device
Heating off
- Switch off the device
- Let the device cool down
- Switch on the device
Heating off
- Switch off the device and switch on again
Heating off
- Switch off the device
- Let the device cool down
- Switch on the device
Heating off
- Switch off the device and switch on again
- Let the device cool down
- Switch on the device
- Check the safety temperature
Heating off
- Switch off the device and switch on again
Heating off
- Let the device cool down
- Fill up the tempering medium
- Switch off the device and switch on again
Heating off
- Switch off the device and switch on again
Heating off
- Switch off the device and switch on again
Heating off
- Switch off the device and switch on again
Heating off
- Switch off the device and switch on again
Accessories
HB 10.2 Cover (incl. spray guard)
remove objects from around the IR
interface
setting
15