Proceed as follows in such cases:
Carry out corrective measures
Restart device
Error message Cause
between RV 10 and HB 10
- IR connection to RV 10 disconnected
E 3
- Temperature inside the device is too high
E 9
- Fault in storing the safety circuit set
temperatures
E 21
- Safety relay does not open
E 23
- Adjustable safety circuit defective
E 24
- Set safety temperature exceeded
faulty. The heater or the supply line is
disconnected.
E 26
- Dry running
E 28
- Breakage of the controller sensor
E 29
- Short-circuit in the safety sensor
E 30
- Short-circuit in the controller sensor
E 32
- Error in calibration
If the actions described fail to resolve the fault or another error code is displayed then take one of the following steps:
- contact the IKA
®
service department,
- send the device for repair, including a short description of the fault.
HB 10.1 Spray guard

Error message

Effect
Heating off
Heating off
Heating off
Heating off
Heating off
Heating off
Heating off
Heating off
Heating off
Heating off

Accessories

Correction
- Clean the IR interface or
remove objects from around the IR
interface
- Switch on the device
- Switch off the device
- Let the device cool down
- Switch on the device
- Switch off the device
- Let the device cool down
- Switch off the device and switch on again
- Switch off the device
- Let the device cool down
- Switch on the device
- Switch off the device and switch on again
- Let the device cool down
- Switch on the device
- Check the safety temperature
setting
- Let the device cool down
- Fill up the tempering medium
- Switch off the device and switch on again
- Switch off the device and switch on again
- Switch off the device and switch on again
- Switch off the device and switch on again
- Switch off the device and switch on again
15