DJO Global ARTROMOT-E2 Mode D'emploi page 45

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 52
6.3.2 ARTROMOT®-E2 compact
The following operating steps must
be completed before transporting the
ARTROMOT®-E2 compact:
1. Activate the "transport setting" func-
tion
in the menu (see also
5.3) and start the ARTROMOT®-E2
compact.
2. Turn off the ARTROMOT®-E2 compact
power switch (19).
3. Remove the power cord (18) and discon-
nect the motion element (9) and the
programming unit (16).
4. Remove pin (26) on the underside and
take off the upper arm support.
5. Remove pin (5) and take off the
swiveling mechanism of the upper arm
support. set the guide of the upper arm
support to the transport setting (paral-
lel with motor A).
6. Grasping the motion element at the
bracket (27) for adjustment of the posi-
tioning angle, open the locking screw (1)
and pull out the motion element.
7. Open the locking screw for adjustment
of the elbow angle adjustment (2) and
separate the oval tube from the motion
element.
8. Open locking screw (24) on the un-
derside of the stand, then remove the
column with the electronics box from
the star-shaped stand.
9. disassemble the stand by loosening the
five legs with the special tool supplied
and remove them.
10. unscrew the locking screw (7) from
motor b (12), remove the entire forearm
support including the slider (4) and
the hand support (13 + 14). Afterwards
screw the locking screw (7) back into
motor b.
11. Only use the original shipping box for
transporting the device. Ormed Gmbh
cannot be held liable for transport dam-
age, if the device was not shipped in its
original shipping box.
82
12. Put the programming unit (15) in the
supplied box, using sufficient padding.
13. Pack the following items in the longer
box provided, again using sufficient
padding:
• upper arm support
• Oval tube of the motion element
• Column with electronics box
• Entire forearm support with slider and
hand support
14. Pack the following items in the shorter
box provided, again using sufficient
padding:
• star-shaped stand
• 5 legs with castors
• 5 screws to attach the legs
• 5 washers to attach the legs
• 1 Allen key
15. Please put the motion element, the
power cord and the three smaller boxes
in the respective cut-outs of the poly-
styrene block.
16. Then close the box.
Warning!
m
shock hazard —
Allow the ARTROMOT®-E2 compact
to reach room temperature before use.
if the device has been transported at
temperatures below 0 °C (32°F), leave
it to dry at room temperature for about
2 hours, until any condensation has
disappeared.
83

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Artromot-e2 compact

Table des Matières