Hach POLYMETRON 9240 Multi-Channel Sodium Manuel D'utilisation De Base page 11

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 51
9. The system then allows you to regulate the next channel until all flow rates have been set for the
configured channels.
10. Once all flow rates for the configured channels have been set, an Action completed message
will be displayed. Select Esc to exit.
Sample pH conditioning check
Non-cationic applications
1. Install a calibrated pH sensor in the center position of the measurement chamber which is
normally used for the ISE sodium electrode (No. 1 in
2. On the analyzer, press Start on the main menu to start the measurement process.
3. For each channel, check that the pH value of the sample after conditioning is greater than 10.5. If
not, check the quality of the conditioning product used.
Cationic applications
1. With a calibrated pH sensor, measure the pH of the sample for each channel outside of the
analyzer.
2. The analyzer measurement sequence will be factory set depending on the number of configured
channels (e.g. 1 2 3 4 * for a 4-channel analyzer or 1 2 * for a 2-channel analyzer). Verify this
sequence is correctly setup (see
3. Set the following measurement timings (see
• CYCLE TIME to 11 min
• ON LINE MEAS to 8 min
• SEARCH STAB to No
4. For each channel, determine the gas injection time ratio depending on the sample pH. Enter this
value into the analyzer as described in
on page 16. The standard values are:
• pH = 2.0 - Tgas/Twater ratio = 180%
• pH = 2.3 - Tgas/Twater ratio = 80%
• pH = 2.6 - Tgas/Twater ratio = 50%
• pH = 2.9 - Tgas/Twater ratio = 30%
• pH = 3.5 - Tgas/Twater ratio = 15%
• pH = 4.0 - Tgas/Twater ratio = 10%
5. Install the same pH sensor in the center position of the measurement chamber (No. 2 in
on page 7).
6. On the analyzer, press Start on the main menu to start the measurement process.
7. For each channel, measure the pH in the conditioned sample to check if the pump ratios are
efficient enough to obtain a pH of around 11.0. If necessary, update the ratio to maintain a final
constant pH of 11.0 ± 0.2.
Reference electrode installation
Figure 1
Sequence
on page 19 for more details).
Measure steps
Total gas/water ratio (cationic applications only)
on page 7).
on page 16):
Figure 1
English 11

Publicité

Table des Matières
loading

Table des Matières