Advanced Operation - Hach sensION+ EC71 Manuel D'utilisation

Masquer les pouces Voir aussi pour sensION+ EC71:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 17
1. Pour the buffers or calibration solutions into the labeled calibration
beakers.
2. From the main menu use the
CALIBRATION parameter. Confirm.
3. If required select the Operator ID (1 to 10) and confirm.
4. Rinse the probe with deionized water and put the probe into the first
calibration beaker. Be sure that there are no air bubbles in the probe
measuring chamber.
5. Push
to Start calibration.
6. Push
to measure the first calibration solution.
The next calibration solution is shown.
7. Rinse the probe with deionized water and put the probe into the
second calibration beaker. Be sure that there are no air bubbles in
the probe measuring chamber.
8. Push
to measure the second calibration solution.
The next calibration solution is shown.
9. Rinse the probe with deionized water and put the probe into the third
calibration beaker. Be sure that there are no air bubbles in the probe
measuring chamber.
10. Push
to measure the third calibration solution.
When the calibration is good, the display briefly shows Calibration
OK and then returns to the main menu.
Note: When a printer is connected the print menu opens and the result can be
printed.
View the calibration data
Data from the most recent calibration can be shown.
1. From the main menu push
2. Use the
to view the last calibration data. The meter returns
automatically to the measurement screen after 3 seconds.
or
and and to select the
.
Set the calibration reminder
The calibration reminder can be set between 0 to 99 days (default
15 days). The display shows the remaining time to the new calibration.
Note: When 0 days is selected, the calibration reminder is turned off.
1. From the main menu use the
Confirm.
2. Use the
to enter the calibration menu.
3. Use the
or
to select Cal. frequency and confirm.
4. Use the and to advance to the next step and use the
change a value. Confirm.
Push
to start the calibration.

Advanced operation

Use a sample ID
The sample ID tag is used to associate readings with a particular sample
location. If assigned, stored data will include this ID.
1. From the main menu use the
2. Use the
or
to select Sample ID and confirm.
3. Use the
or
to select
Option
Description
Automatic A consecutive number will be automatically assigned to every
sample.
Manual
A keyboard or a barcode scanner is required to enter the
sample ID name before taking a measurement (maximum
15 characters).
Select the data output
Data can be stored or transferred to a printer or to a PC.
1. From the main menu use the
2. Use the
or
to select Data Output and confirm.
or
to select CALIBRATION.
or
or
to select SYSTEM. Confirm.
or
to select SYSTEM. Confirm.
English 9
to

Publicité

Table des Matières
loading

Table des Matières