Sprint / Stroller / Hi Flow Stroller
Caution: Check liquid level gauge only
after the valve is closed.
Caution: If the Sprint/Stroller/Hi Flow Stroller is
being refilled immediately after a period of use,
fill it, and then wait 10 minutes after filling with
the FCV set at zero for the pressure to stabilize.
Then operate as normal.
WARNING: DO NOT OPERATE SPRINT/
STROLLER/HI FLOW STROLLER WHILE
ATTACHED TO THE STATIONARY UNIT.
WARNING: IF THE SPRINT/STROLLER/HI FLOW
STROLLER DOES NOT SEPARATE EASILY, DO
NOT USE FORCE. THE UNITS MAY BE FROZEN
TOGETHER. LEAVE THE UNITS CONNECTED AND
WAIT UNTIL THE UNITS WARM UP — THEN
THEY WILL SEPARATE EASILY. DO NOT TOUCH
ANY FROSTED PARTS. WARNING: SHOULD THE
LEAKAGE BE EXCESSIVE TO THE POINT THAT
A STREAM OF LIQUID IS PRESENT, LEAVE THE
AREA AND CALL YOUR HEALTH CARE PROVIDER
IMMEDIATELY.
WARNING: LIQUID OR COLD GASEOUS OXYGEN
CAN FROSTBITE THE SKIN. OXYGEN WILL RESULT
IN INCREASED FIRE HAZARD.
Caution: Should there be any liquid
leakage from the stationary unit
or Sprint/Stroller/Hi Flow Stroller
separating the units, set the Sprint/
Stroller/Hi Flow Stroller aside, ensuring
it remains vertical, leave the room,
and call your health care provider
immediately.
Note: To Equipment Provider: The following oxygen
administration accessories are recommended for use
with the Sprint / Stroller / Hi Flow Stroller:
• Nasal Cannula: CAIRE Part Number 5408-SEQ
• Firebreak: CAIRE Part Number 21126636
A firebreak is recommended for use with any can-
nula.
• CAIRE offers a firebreak intended to be used in
conjunction with the oxygen portable. The firebreak
is a thermal fuse to stop the flow of gas in the event
that the downstream cannula or oxygen tubing is
ignited and burns to the firebreak. It is placed in-line
with the nasal cannula or oxygen tubing between
the patient and the oxygen outlet of the Sprint
/ Stroller / Hi Flow Stroller. For proper use of the
firebreak, always refer to the manufacturer's instruc-
tions (included with each firebreak kit).
• For any additional recommended accessories,
please see the Accessories Catalog (PN MLLOX0010)
available on www.caireinc.com.
8 - ENG
PN MN236-C4 C | User Manual
Operating Instructions
(Cannula is not included.)
1. Push the cannula breathing tube firmly on to the
oxygen tube connector.
2. Adjust your cannula breathing tube to the proper
position so that you will be able to breathe the
oxygen comfortably.
3. Turn the flow control knob on top of the Sprint/
Stroller/Hi Flow Stroller clockwise until the
prescribed flow rate (numeral) is visible in the
knob "window" and a positive detent is felt.
Caution: Out-of-specification oxygen
flow will result if the flow control knob
is set between flow rates.
4. You should be receiving oxygen now. To stop
flow of oxygen, turn the flow control knob
counterclockwise to the off (0) position.
Caution: Always turn the flow control
knob off (0) when not in use.